Z-2174-2017Class II
Endopath ETS-FLEX Endoscopic Articulating Linear Cutter, VASCULAR/THIN 35MM
Ethicon Endo-Surgery Inc / initiated 2017-03-31 / Terminated / root cause: manufacturing process
Ethicon Endo-Surgery Inc began a recall of Endopath ETS-FLEX Endoscopic Articulating Linear Cutter, VASCULAR/THIN 35MM on . FDA classified it as Class II and gives the reason as "There is a risk that the pinion gear in the device could fail under extreme use cases". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2174-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ethicon Endo-Surgery Inc
- Product
- Endopath ETS-FLEX Endoscopic Articulating Linear Cutter, VASCULAR/THIN 35MM
- Product code
- GDW Staple, Implantable
- Reason for recall
- There is a risk that the pinion gear in the device could fail under extreme use cases.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Consignees were sent notification letters and were asked to complete the Business Reply Form (BRF).
- Quantity in commerce
- 792 units
- Distribution
- Worldwide distribution. The previously recalled products were distributed to the following countries: China, Japan, Singapore, and South Korea.
- Product codes and lots
- N91U0W, N91U0X, N91L2P, N91N1J, and N91N2G