Z-1877-2017Class II

ViperWire Advance with Flextip Peripheral Atherectomy Guide Wire, Part Number: 7-10026-01,Model Number...

Cardiovascular Systems Incorporated / initiated 2017-02-09 / Terminated / root cause: manufacturing process

Cardiovascular Systems Inc began a recall of ViperWire Advance with Flextip Peripheral Atherectomy Guide Wire, Part Number: 7-10026-01,Model Number: VPR-GW-FT14. Product Usage: The Stealth 360 Orbital PAD... on . FDA classified it as Class II and gives the reason as "CSI is recalling three lots of VIPERWIRE ADVANCE WITH FLEXTIP PERIPHERAL ATHERECTOMY GUIDEWIRE . It was discovered that 0.018in ViperWire Advance with Flextip product was packaged as a 0.014in ViperWire Advance with...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1877-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cardiovascular Systems Incorporated
Product
ViperWire Advance with Flextip Peripheral Atherectomy Guide Wire, Part Number: 7-10026-01,Model Number: VPR-GW-FT14. Product Usage: The Stealth 360 Orbital PAD System and the Diamondback 360 Peripheral Orbital Atherectomy System are percutaneous orbital atherectomy systems indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy. The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The OAS is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.
Product code
MCW Catheter, Peripheral, Atherectomy
Reason for recall
CSI is recalling three lots of VIPERWIRE ADVANCE WITH FLEXTIP PERIPHERAL ATHERECTOMY GUIDEWIRE . It was discovered that 0.018in ViperWire Advance with Flextip product was packaged as a 0.014in ViperWire Advance with Flextip.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On 2/9/2017, consignees were sent a CSI "Urgent Medical Device Recall" letter dated 09 February 2017. The letter described the Affected Product, Recall Description, & Instructions. Advised consignees to immediately remove, return the product & complete the Customer Acknowledgement Form. For further information contact Customer Service, Cardiovascular Systems, Inc., 1225 Old Highway 8 NW Saint Paul, MN 55112 877-274-0901 - Tel, 612-677-3355 - Fax.
Quantity in commerce
595
Distribution
US Nationwide in the states of: AK, AL, AR, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, NC, NJ, NY, OH, OR, PA, SC, SD, TN, TX, and WA
Product codes and lots
Lot #: 175069, 175070, 176255

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1877-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1877-2017Class IITerminatedVoluntary: Firm initiated