Destination Renal Guiding Sheath Intended for the introduction of interventional and diagnostic devices into...
Terumo Medical Corporation began a recall of Destination Renal Guiding Sheath Intended for the introduction of interventional and diagnostic devices into the human vasculature on . FDA classified it as Class II and gives the reason as "Based on internal investigation and testing that revealed the distal end of certain lots of Terumo Medical Destination products may not contain the labeled 5 cm of coating. While there have been no complaints or patient...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2033-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Medical Corp.
- Product
- Destination Renal Guiding Sheath Intended for the introduction of interventional and diagnostic devices into the human vasculature.
- Product code
- DYB Introducer, Catheter
- Reason for recall
- Based on internal investigation and testing that revealed the distal end of certain lots of Terumo Medical Destination products may not contain the labeled 5 cm of coating. While there have been no complaints or patient related incidents reported, the lack of coating could render it difficult to navigate the device to the target vasculature.
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- An Urgent Product Bulletin Recall of Terumo Medical Product letter, dated November 6, 2015, was sent to customers. The letter identified the affected product and the reason for the voluntary recall. The letter asked customers to immediately discontinue use and quarantine any of the affected product listed. Customers are to complete the enclosed Contact & Inventory Confirmation Form and e-mail [email protected] or fax to 732-412-4121, Attn: Destination Recall. For questions or further information, customers can contact Customer Care at 800-888-3786 or their local sales representative.
- Quantity in commerce
- 5,252 eaches
- Distribution
- Nationwide Distribution.
- Product codes and lots
- Terumo Product Code - RSR01 Affected Lot Numbers RN19 RN26 RP03
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2033-2017 | Class II | Terminated | Voluntary: Firm initiated |