Z-2033-2017Class II

Destination Renal Guiding Sheath Intended for the introduction of interventional and diagnostic devices into...

Terumo Medical Corp. / initiated 2015-11-06 / Terminated / root cause: labeling and UDI

Terumo Medical Corporation began a recall of Destination Renal Guiding Sheath Intended for the introduction of interventional and diagnostic devices into the human vasculature on . FDA classified it as Class II and gives the reason as "Based on internal investigation and testing that revealed the distal end of certain lots of Terumo Medical Destination products may not contain the labeled 5 cm of coating. While there have been no complaints or patient...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2033-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Medical Corp.
Product
Destination Renal Guiding Sheath Intended for the introduction of interventional and diagnostic devices into the human vasculature.
Product code
DYB Introducer, Catheter
Reason for recall
Based on internal investigation and testing that revealed the distal end of certain lots of Terumo Medical Destination products may not contain the labeled 5 cm of coating. While there have been no complaints or patient related incidents reported, the lack of coating could render it difficult to navigate the device to the target vasculature.
Root cause tag
labeling and UDI
FDA root cause
Other
Action
An Urgent Product Bulletin Recall of Terumo Medical Product letter, dated November 6, 2015, was sent to customers. The letter identified the affected product and the reason for the voluntary recall. The letter asked customers to immediately discontinue use and quarantine any of the affected product listed. Customers are to complete the enclosed Contact & Inventory Confirmation Form and e-mail [email protected] or fax to 732-412-4121, Attn: Destination Recall. For questions or further information, customers can contact Customer Care at 800-888-3786 or their local sales representative.
Quantity in commerce
5,252 eaches
Distribution
Nationwide Distribution.
Product codes and lots
Terumo Product Code - RSR01  Affected Lot Numbers  RN19  RN26  RP03

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2033-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2033-2017Class IITerminatedVoluntary: Firm initiated