Z-2164-2017Class II

SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK Product Code/Part...

Datascope Corp. / initiated 2017-04-12 / Terminated / root cause: design

Datascope Corporation began a recall of SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK Product Code/Part Number: 0684-00-0576-01U is a cardiac assist device... on . FDA classified it as Class II and gives the reason as "Maquet/Getinge is initiating a voluntary product removal involving four lot numbers of SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK since during internal pressure (leak) testing...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2164-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Datascope Corp.
Product
SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK Product Code/Part Number: 0684-00-0576-01U is a cardiac assist device which support the heart's left ventricle by increasing coronary perfusion and reducing the work of the left ventricle.
Product code
DSP System, Balloon, Intra-Aortic And Control
Reason for recall
Maquet/Getinge is initiating a voluntary product removal involving four lot numbers of SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK since during internal pressure (leak) testing of the SENSATION PLUS 8Fr. 50cc, IABC, failures related to the tip seal leak were identified.
Root cause tag
design
FDA root cause
Device Design
Action
Getinge notified their customers with a recall letter/return response form via Fed Ex on 4/12/2017. Getinge notified their customers again through an expansion on 10/16/2017.
Quantity in commerce
996 UNITS
Distribution
Nationwide
Product codes and lots
300004515,  3000043376 3000040272 3000038372  added lot numbers:  3000043374 and 3000045513

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2164-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2164-2017Class IITerminatedVoluntary: Firm initiated