SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK Product Code/Part...
Datascope Corporation began a recall of SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK Product Code/Part Number: 0684-00-0576-01U is a cardiac assist device... on . FDA classified it as Class II and gives the reason as "Maquet/Getinge is initiating a voluntary product removal involving four lot numbers of SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK since during internal pressure (leak) testing...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2164-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Datascope Corp.
- Product
- SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK Product Code/Part Number: 0684-00-0576-01U is a cardiac assist device which support the heart's left ventricle by increasing coronary perfusion and reducing the work of the left ventricle.
- Product code
- DSP System, Balloon, Intra-Aortic And Control
- Reason for recall
- Maquet/Getinge is initiating a voluntary product removal involving four lot numbers of SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK since during internal pressure (leak) testing of the SENSATION PLUS 8Fr. 50cc, IABC, failures related to the tip seal leak were identified.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Getinge notified their customers with a recall letter/return response form via Fed Ex on 4/12/2017. Getinge notified their customers again through an expansion on 10/16/2017.
- Quantity in commerce
- 996 UNITS
- Distribution
- Nationwide
- Product codes and lots
- 300004515, 3000043376 3000040272 3000038372 added lot numbers: 3000043374 and 3000045513
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2164-2017 | Class II | Terminated | Voluntary: Firm initiated |