Z-1923-2017Class II

Schiller AG ISA CO2 Sensor (AKA: PhaseIn Unit), Merge Part Number HW-HEMO-00001. The PhaseIn Unit is an...

Merge Healthcare Incorporated / initiated 2015-03-30 / Terminated

Merge Healthcare, Inc. began a recall of Schiller AG ISA CO2 Sensor (AKA: PhaseIn Unit), Merge Part Number HW-HEMO-00001. The PhaseIn Unit is an optional accessory (sold separately) to the Merge Hemo... on . FDA classified it as Class II and gives the reason as "The units may display the alarm "sampling line clogged" after the flow zeroing procedure that is conducted automatically 1 hour after start up. This could result in the potential of a small bias added to the CO2...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1923-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merge Healthcare Incorporated
Product
Schiller AG ISA CO2 Sensor (AKA: PhaseIn Unit), Merge Part Number HW-HEMO-00001. The PhaseIn Unit is an optional accessory (sold separately) to the Merge Hemo Patient Data Module (PDM). The PDM is a major component of the Merge Hemo system.
Product code
DQK Computer, Diagnostic, Programmable
Reason for recall
The units may display the alarm "sampling line clogged" after the flow zeroing procedure that is conducted automatically 1 hour after start up. This could result in the potential of a small bias added to the CO2 measurement and affect the alarm notification.
FDA root cause
Unknown/Undetermined by firm
Action
The recall was initiated by telephone on March 30, 2015. Follow-up phone calls were made July 6, 2015. The firm was unable to provide a telephone script. The recalling firm also issued letters dated March 28, 2016, via email on April 6, 2016, dated July 11, 2016. The letters contained the same content, notified the customer of the issue, and informed them a fix was available. The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. The notification also informs the customer their response is required using the enclosed form and the return addressed envelope. The response form asks, in part, if they have the affected ISA CO2 sensors at their facility and are they interested in accepting the fix. If they decline the fix, they are asked to provide the reason why. For further questions, please call (877) 741-5369.
Quantity in commerce
23 sensors (PhaseIn Units)
Distribution
US Distribution was made to medical facilities located in AL, FL, ID, LA, MO, NC, OK, SC, TX, and UT. There was government distribution and no foreign/military distribution.
Product codes and lots
ISA CO2 Sensor serial numbers:  107349, 107357, 107358, 107363, 107370, 107377, 107390, 107407, 107435,  107476, 107509, 107510, 107556, 107659, 107700, 107707, 107768, 107776,  107831, 107832, 107836, 107839, and 107841.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1923-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1923-2017Class IITerminatedVoluntary: Firm initiated