Z-1922-2017Class II

The JETSTREAM is a rotational atherectomy catheter system designed for use in debulking and treating vascular...

Boston Scientific Corporation / initiated 2017-04-05 / Terminated / root cause: manufacturing process

Boston Scientific Corporation began a recall of The JETSTREAM is a rotational atherectomy catheter system designed for use in debulking and treating vascular disease in the peripheral vasculature. The System... on . FDA classified it as Class II and gives the reason as "Boston Scientific is initiating a voluntary removal of two lots of JETSTREAM SC 1.85mm and one lot of JETSTREAM XC 2.1mm Atherectomy Catheters due to these units being mislabeled. Some 1.85mm JETSTREAM SC catheters may...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1922-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
The JETSTREAM is a rotational atherectomy catheter system designed for use in debulking and treating vascular disease in the peripheral vasculature. The System includes multiple distal ports located at the catheter tip, which are designed to provide independent infusion and aspiration functions for the active removal of fluid, excised tissue and thrombus from the peripheral treatment site.
Product code
MCW Catheter, Peripheral, Atherectomy
Reason for recall
Boston Scientific is initiating a voluntary removal of two lots of JETSTREAM SC 1.85mm and one lot of JETSTREAM XC 2.1mm Atherectomy Catheters due to these units being mislabeled. Some 1.85mm JETSTREAM SC catheters may have been labeled as 2.1mm catheters, and some 2.1mm JETSTREAM XC catheters may have been labeled as 1.85mm catheters.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Consignees were sent a Boston Scientific "Urgent Medical Device Removal" letter dated April 5, 2017. The letter was addressed to Materials Manager / Field Action Contact. The letter described the issue and the prduct involved in the recall. Advised consignees to segregate and return the affected product to Boston Scintific. This recall is to be carried to the customer level. Requested consginees to complete and return the Account Reply Verification Tracking Form. Contact your local Sales Representative can answer any questions that you may have regarding this Removal.
Quantity in commerce
137
Distribution
US: AL, AZ, CT, FL, HI, IL IN, IA, KY, LA, ME, MI, MO, NJ,NY, OH, OR, PA, SC, TN, TX, VA, WA, WI. OUS; GERMANY, GREAT BRITAIN, POLAND, NETHERLANDS
Product codes and lots
18775553, 18775552, 18822086

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1922-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1922-2017Class IITerminatedVoluntary: Firm initiated