AQUIOS CL Flow Cytometer, Ref no. B30166, Software Versions 2.0 and 2.0.1
Beckman Coulter Inc. began a recall of AQUIOS CL Flow Cytometer, Ref no. B30166, Software Versions 2.0 and 2.0.1 Product Usage: The AQUIOS CL Flow Cytometer is intended for use with in vitro... on . FDA classified it as Class II and gives the reason as "Beckman Coulter has confirmed that the Export feature located in the Results area of the AQUIOS System software is not correctly exporting Tetra Combo CRD files". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2035-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- AQUIOS CL Flow Cytometer, Ref no. B30166, Software Versions 2.0 and 2.0.1 Product Usage: The AQUIOS CL Flow Cytometer is intended for use with in vitro diagnostic flow cytometric applications using up to four fluorescent detection channels using a blue (488 nm) laser, two light scatter detection channels and electronic volume (EV).
- Product code
- OYE Flow Cytometric Reagents And Accessories.
- Reason for recall
- Beckman Coulter has confirmed that the Export feature located in the Results area of the AQUIOS System software is not correctly exporting Tetra Combo CRD files.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- An Urgent Medical Device Recall letter was sent on 4/3/17 to inform customers that area of the AQUIOS System software is not correctly exporting Tetra Combo CRD files. The letter informs the customers of the impact, action, and resolution. Customers are instructed to complete and return the enclosed response form within 10 days. Customers with questions are instructed to contact Customer Support Center via website, http://www.beckmancoulter.com/customersupport/support, Via phone, call 1-800-369-0333 in the US. Customers outside the US are instructed to contact their local Beckman Coulter Representative.
- Quantity in commerce
- 188 units total (24 units in US)
- Distribution
- Worldwide - US Nationwide in the countries of: Angola, Australia, Belgium, Botswana, Canada, Chile, Columbia, Czech Republic, Denmark, Finland, France, Germany, Ghana, Israel, Italy, Malaysia, Nambia, Netherlands, Pakistan, Panama, Poland, Portugal, Romania, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkey, Uganda, United Kingdom and Zambia
- Product codes and lots
- All
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2035-2017 | Class II | Terminated | Voluntary: Firm initiated |