Dale Medical Stabilock Endotracheal Tube Holder Product # 270 The Dale Stabilock Endotracheal Tube Bolder 270...
Dale Medical Products, Inc. began a recall of Dale Medical Stabilock Endotracheal Tube Holder Product # 270 The Dale Stabilock Endotracheal Tube Bolder 270 is a 2 piece device consisting of an adhesive... on . FDA classified it as Class II and gives the reason as "ET Tube Holders contain an adhesive faceplate that may lose its adhesive properties and will not adhere to the patient's face". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1927-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Dale Medical Products, Inc.
- Product
- Dale Medical Stabilock Endotracheal Tube Holder Product # 270 The Dale Stabilock Endotracheal Tube Bolder 270 is a 2 piece device consisting of an adhesive base, Dale product number 273, that attaches to the patient's skin and a cushioned neckband with tube channel that attaches with hook & loop closures. The 270 is used to secure and stabilize endotracheal tubes, preventing unnecessary movement which can help prevent accidental .extubation
- Product code
- CBH Device, Fixation, Tracheal Tube
- Reason for recall
- ET Tube Holders contain an adhesive faceplate that may lose its adhesive properties and will not adhere to the patient's face.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Dale Medical notified accounts (OUS) on March 24, 2017 and US accounts on April 7, 2017 advising them to review inventory, if you have any of the noted Lot numbers in stock contact your customers that may have received the identified lot numbers of 270 ET Tube holders and remove suspected product.. Dale Medical will replace product. For questions customers were instructed to call 508-695-9316 ext. 143
- Quantity in commerce
- 34,770 total
- Distribution
- Worldwide Distribution -US ( Nationwide) and Internationally to Canada, Costa Rica, Norway, Paraguay, Ireland, Panama, Ecuador, Hong Kong, Malaysia, Italy, UAE, Guatemala, Iceland, Indonesia, and Saudi Arabia.
- Product codes and lots
- Lot Numbers: 02 23 16, 04 11 16, 05 11 16, 02 24 16 , 04 19 16 , 05 13 16, 03 04 16, 04 20 16 , 05 16 16, 03 07 16, 04 21 16, 05 18 16, 03 08 16, 05 03 16 , 05 20 16, 03 10 16 , 05 05 16, 05 24 16, 03 11 16 , 05 06 16 , 05 26 16, 04 07 16, 05 09 16 , 05 31 16, 04 08 16, 05 10 16 , 06 02 16
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1927-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1928-2017 | Class II | Terminated | Voluntary: Firm initiated |