SpO2 connecting cable for Masimo LNCS sensor (Merge Part Number: HW-HEMO-00014)
Merge Healthcare, Inc. began a recall of SpO2 connecting cable for Masimo LNCS sensor (Merge Part Number: HW-HEMO-00014), Product Usage: Merge Hemodynamics monitors, measures, and records physiologic... on . FDA classified it as Class II and gives the reason as "Under certain conditions, the cables used with the Hemo system during a patient procedure could stop displaying the patient SpO2 causing a slight delay in patient care". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2640-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Merge Healthcare Incorporated
- Product
- SpO2 connecting cable for Masimo LNCS sensor (Merge Part Number: HW-HEMO-00014), Product Usage: Merge Hemodynamics monitors, measures, and records physiologic data from a human patient undergoing a cardiac catheterization procedure. The Monitoring System is for the monitoring of vital parameters including ECG, SpO2, invasive blood pressure, temperature, NIBP and C02, and for the evaluation of resting ECG, arrhythmias, ST-segments and cardiac output.
- Product code
- DQK Computer, Diagnostic, Programmable
- Reason for recall
- Under certain conditions, the cables used with the Hemo system during a patient procedure could stop displaying the patient SpO2 causing a slight delay in patient care.
- FDA root cause
- Component change control
- Action
- Merge sent an Urgent: Medical Device Recall letter dated March 28, 2016 via e-mail the week of 4/4/2016. If no email address was available, the letters were issued via certified mail. A second letter dated July 11, 2016 was issued to non-responders. The letter identified the affected product problem and actions to be taken.
- Quantity in commerce
- 219 cables
- Distribution
- Distribution was nationwide. There was also government and military distribution. There was no foreign distribution.
- Product codes and lots
- The following lot numbers are affected and are not specific to each cable/sensor type: 09AP8, 10AEU, 10KBN, 09CF9, 10BMN, 10MDN, 09CNS, 10FEN, 10MDN, 09KG6, 10F95, 10DUT, 09KG6, 10F95, 10DDK, 09NMN, 10BMN, 10DUT, 10A81, 10GJB, 10KBS, 09NMN, 10BMN, 10DUT, 10AEU, 10GJB, 10KBS, 10AEU, 10KBN, 10KBS, 11AGS, 11KAJ, 11CFT, 10KFP, 11N61, 11DAY, 11AGS, 11NG2, 11CFT, 11AGS, 12CNX, 11DDK, 12FMH, 12C68, 11FBS, 12GEB, 12C68, 11JBK, 12GEB, and 12FMH.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2640-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2641-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2642-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2643-2017 | Class II | Terminated | Voluntary: Firm initiated |