Z-2662-2017Class II
Logic Fit Tibial Tamp Head
Exactech, Inc. / initiated 2017-05-25 / Terminated / root cause: labeling and UDI
Exactech, Inc. began a recall of Logic Fit Tibial Tamp Head on . FDA classified it as Class II and gives the reason as "The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to remove the tamp assembly from the bone, i.e., to include a caution statement about the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2662-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Exactech, Inc.
- Product
- Logic Fit Tibial Tamp Head
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to remove the tamp assembly from the bone, i.e., to include a caution statement about the potential for instrument breakage if the tamp handle/guide is misused by impacting in retro-grade.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Consignees were sent a Field Advisory Notice dated 5/22/2017 by mail informing them updated surgical techniques are available on Exactech's website (www.exac.com). The notice requested consignees: Extend the information to accounts in possession of the product Complete & return the attached Field Advisory Notice Response Form to Exactech
- Quantity in commerce
- 573
- Distribution
- AL, AZ, CA, CA, CO, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, ME, MN, NC, NJ, NV, NY, OH, OK, OR, RI, SC, TN, TX, VA, WA, WI, Hawaii and Puerto Rico Australia, France, Germany, India, Italy, Japan, Korea, Luxembourg, Singapore, Spain, Switzerland, The Netherlands and United Kingdom
- Product codes and lots
- Catalog #: 213-75-01 Lot #'s: 37166-002, 44944-004, 44944-005, 44944-006, 48073-010, 48073-011, 48073-012, 49036-017, 51180-004, 51180-005, 51180-006, 51180-008, 51754-031, 51754-032, 51754-033, 51754-034, 59743-001, 61872-001,and 61872-002.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2662-2017 | Class II | Terminated | Voluntary: Firm initiated |