Exactech 1.5" Novation Calcar Planer Guide Tip To assist the surgeon in the implantation of Novation hip...
Exactech, Inc. began a recall of Exactech 1.5" Novation Calcar Planer Guide Tip To assist the surgeon in the implantation of Novation hip system components according to a conventional... on . FDA classified it as Class II and gives the reason as "The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to perform calcar planing to the bone, i.e., to include a caution statement about the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2644-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Exactech, Inc.
- Product
- Exactech 1.5" Novation Calcar Planer Guide Tip To assist the surgeon in the implantation of Novation hip system components according to a conventional technique for total hip replacement. Specifically, these instruments are reusable, surgically invasive, and intended for transient use. ech 1.5" Novation Calcar Planer Guide Tip
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to perform calcar planing to the bone, i.e., to include a caution statement about the potential for instrument breakage if bending or torsional forces are applied.
- Root cause tag
- labeling and UDI
- FDA root cause
- Under Investigation by firm
- Action
- Exactech sent an Field Advisory Notice dated May 22, 2017, by mail informing the customers the updated surgical technique is available on Exactech's website (www.exac.com). The notice requested consignees: Extend the information to accounts in possession of the product Complete & return the attached Field Advisory Notice Response Form to Exactech. For further questions, please call (800) 392-2832.
- Quantity in commerce
- 477
- Distribution
- Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and to the states of : AL, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, ME, MN, NC, NJ, NV, NY, OH, OK, OR, RI, SC, TN, TX, VA, WA, WI, Hawaii., and to the countries of : Argentina, Australia, Austria, Brazil, China, Colombia, France, Germany, Guatemala, India, Italy, Japan, Spain, Switzerland, Turkey, and United Kingdom
- Product codes and lots
- Catalog #: 161-30-00 with the manufacturing lots #'s 38878-001, 38878-002, 49777-001, 51825-005, 51825-006, 52224-007, 51825-006, 52224-005, 52224-006, 57167-001, 59775-003, 62580-004, 66464-003, 66464-004, 70720-001, 71530-001, 73477-002, 75153-001, and 74999-001.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2644-2017 | Class II | Terminated | Voluntary: Firm initiated |