Z-2093-2017Class II

Spacelabs Healthcare Arkon Anesthesia Delivery System is intended for use in the hospital environment and...

GE Healthcare / initiated 2017-04-14 / Terminated / root cause: design

Del Mar Reynolds Medical, Ltd. began a recall of Spacelabs Healthcare Arkon Anesthesia Delivery System is intended for use in the hospital environment and operating room. Absorber Adapter Assembly Service... on . FDA classified it as Class II and gives the reason as "Reports of the Canister Carrier breaking during absorbent exchange or during anesthesia machine transit, causing a leak in the breathing circuit, which could cause hypoventilation in the patient". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2093-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Spacelabs Healthcare Arkon Anesthesia Delivery System is intended for use in the hospital environment and operating room. Absorber Adapter Assembly Service Kit, part number: 050-9045-00. Absorber Adapter Assembly, part number: 650-1648-01. Arkon Absorber Canister Carrier, part number: 380-0916-01.
Product code
BSZ Gas-Machine, Anesthesia
Reason for recall
Reports of the Canister Carrier breaking during absorbent exchange or during anesthesia machine transit, causing a leak in the breathing circuit, which could cause hypoventilation in the patient.
Root cause tag
design
FDA root cause
Device Design
Action
Spacelabs sent an URGENT - MEDICAL DEVICE CORRECTION letter dated April 13, 2017, to US customers via priority service. The customer letter (translated as necessary) will be emailed to all international subsidiaries and distributors of record on April 14, 2017. The letter announced that the The Arkon Anesthesia Delivery System, Model 99999, and the Absorber Adapter Assembly Kit part number 050-9045-00, and Absorber Adapter Assembly part number 650-1648-01 are being recalled. Customers are advised that The Arkon Anesthesia Delivery Systems Model 99999 with Absorber Adapter Assembly part number 650-1648-00 is NOT affected by this recall. Spacelabs Healthcare is developing a hardware upgrade to improve the robustness of the affected Absorber Adapter Assembly. A Spacelabs representative will contact consignees once the upgrade is available to schedule a convenient time to update their Arkon Anesthesia Delivery System(s) at no cost. For additional information or technical assistance, please contact: 1-800-522-7025 select 4 for Anesthesia Technical Support
Quantity in commerce
62 total (US= 19 devices and 32 kits; International = 11 devices)
Distribution
US distribution in FL, LA, MS, NC, NC, SC and WY and Internationally to Panama and United Kingdom.
Product codes and lots
*********the serial numbers and part numbers of units distributed in the US *********  Absorber Adapter Assembly Service Kit part number: 050-9045-00.  Devices serial numbers: ARKN-000395, ARKN-000396, ARKN-000397, ARKN-000398, ARKN-000399, ARKN-000400, ARKN-000401, ARKN-000402, ARKN-000403, ARKN-000404, ARKN-000405, ARKN-000406, ARKN-000407, ARKN-000408, ARKN-000409, ARKN-000410, ARKN-000411, ARKN-000412, and ARKN-000413.  *********the serial numbers of units distributed international **************************  ARKN-000375, ARKN-000414, ARKN-000415, ARKN-000416, ARKN-000417, ARKN-000418, ARKN-000419, ARKN-000420, ARKN-000421, ARKN-000422, and ARKN-000423.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2093-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2093-2017Class IITerminatedVoluntary: Firm initiated