NDI P7 Position Sensor, Stereotaxic Instrument
Orthosoft, Inc. dba Zimmer CAS began a recall of NDI P7 Position Sensor, Stereotaxic Instrument on . FDA classified it as Class II and gives the reason as "Zimmer CAS voluntarily conducted a retrospective recall on specific P7 Position Sensors (Cameras), installed on Sesamoid and Sesamoid Plasty CAS workstations due to a series of these components determined to be affected...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2048-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer CAS
- Product
- NDI P7 Position Sensor, Stereotaxic Instrument
- Product code
- HAW Neurological Stereotaxic Instrument
- Reason for recall
- Zimmer CAS voluntarily conducted a retrospective recall on specific P7 Position Sensors (Cameras), installed on Sesamoid and Sesamoid Plasty CAS workstations due to a series of these components determined to be affected with a manufacturing issue whereby they may potentially stop functioning during usage.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Zimmer CAS sent an Field Recall Notice dated June 9, 2009. Zimmer CAS conducted a retrospective recall/Field Correction on given P7 Position Sensors (Cameras). These were installed on given Sesamoid and Sesamoid Plasty CAS workstations. A series of these components have been determined to be affected with a manufacturing issue whereby they may potentially stop functioning during usage. The failure mode consists in the camera stopping to track and interrupting the application with an error code 0X00000020 being displayed on the navigation panel. For any questions or concerns please contact the Zimmer CAS customer service, 1-866-336-7846.
- Quantity in commerce
- 34
- Distribution
- Worldwide Distribution - US Distribution to the states of : KS, MS, WI, NJ, KS, IN, LA, FL and WA., and to the countries of : Canada, Australia, Singapore, Netherlands, UK, Austria, Germany, Switzerland and Czech Republic.
- Product codes and lots
- Part Number: NDI P7 Position Sensor Product Number 201.053 (a component of part number 521.035, Sesamoid Plasty Camera Arm Kit) Lot/Serial Number: 38 units affected: P7-00321, P7-00339, P7-00360, P7-00361, P7-00362, P7-00363, P7-00364, P7-00365, P7-00366, P7-00368, P7-00369, P7-00370, P7-00371, P7-00372, P7-00373, P7-00375, P7-00387, P7-00412, P7-00414, P7-00415, P7-00416, P7-00417, P7-00418, P7-00419, P7-00420, P7-00421, P7-00422, P7-00423, P7-00431, P7-00451, P7-00452, P7-00453, P7-00562, P7-00563, P7-00565, P7-00568, P7-00569, P7-00590 (The lot number and the serial number are the same)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2048-2017 | Class II | Terminated | Voluntary: Firm initiated |