Z-2048-2017Class II

NDI P7 Position Sensor, Stereotaxic Instrument

Zimmer CAS / initiated 2009-06-09 / Terminated / root cause: manufacturing process

Orthosoft, Inc. dba Zimmer CAS began a recall of NDI P7 Position Sensor, Stereotaxic Instrument on . FDA classified it as Class II and gives the reason as "Zimmer CAS voluntarily conducted a retrospective recall on specific P7 Position Sensors (Cameras), installed on Sesamoid and Sesamoid Plasty CAS workstations due to a series of these components determined to be affected...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2048-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer CAS
Product
NDI P7 Position Sensor, Stereotaxic Instrument
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
Zimmer CAS voluntarily conducted a retrospective recall on specific P7 Position Sensors (Cameras), installed on Sesamoid and Sesamoid Plasty CAS workstations due to a series of these components determined to be affected with a manufacturing issue whereby they may potentially stop functioning during usage.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Zimmer CAS sent an Field Recall Notice dated June 9, 2009. Zimmer CAS conducted a retrospective recall/Field Correction on given P7 Position Sensors (Cameras). These were installed on given Sesamoid and Sesamoid Plasty CAS workstations. A series of these components have been determined to be affected with a manufacturing issue whereby they may potentially stop functioning during usage. The failure mode consists in the camera stopping to track and interrupting the application with an error code 0X00000020 being displayed on the navigation panel. For any questions or concerns please contact the Zimmer CAS customer service, 1-866-336-7846.
Quantity in commerce
34
Distribution
Worldwide Distribution - US Distribution to the states of : KS, MS, WI, NJ, KS, IN, LA, FL and WA., and to the countries of : Canada, Australia, Singapore, Netherlands, UK, Austria, Germany, Switzerland and Czech Republic.
Product codes and lots
Part Number: NDI P7 Position Sensor Product Number 201.053 (a component of part number 521.035, Sesamoid Plasty Camera Arm Kit) Lot/Serial Number: 38 units affected: P7-00321, P7-00339, P7-00360, P7-00361, P7-00362, P7-00363, P7-00364, P7-00365, P7-00366, P7-00368, P7-00369, P7-00370, P7-00371, P7-00372, P7-00373, P7-00375, P7-00387, P7-00412, P7-00414, P7-00415, P7-00416, P7-00417, P7-00418, P7-00419, P7-00420, P7-00421, P7-00422, P7-00423, P7-00431, P7-00451, P7-00452, P7-00453, P7-00562, P7-00563, P7-00565, P7-00568, P7-00569, P7-00590 (The lot number and the serial number are the same)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2048-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2048-2017Class IITerminatedVoluntary: Firm initiated