Z-2060-2017Class II

Savaria SL-1000 Stairlift

Savaria Concord Lifts, Inc. / initiated 2017-04-12 / Terminated / root cause: design

Savaria Concord Lifts, Inc. began a recall of Savaria SL-1000 Stairlift on . FDA classified it as Class II and gives the reason as "The electrical controller, combined with motor, causes feedback, damaging a component on the control board, starting a fire". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2060-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Savaria Concord Lifts, Inc.
Product
Savaria SL-1000 Stairlift
Product code
PCD Powered Stairway Chair Lift
Reason for recall
The electrical controller, combined with motor, causes feedback, damaging a component on the control board, starting a fire.
Root cause tag
design
FDA root cause
Component design/selection
Action
The firm sent a letter to dealers dated July 26, 2016. The firm will contact consignees (distributors) via e-mail and phone.
Quantity in commerce
1425 in the United States 734 in Canada
Distribution
Distribution in the United States and Canada
Product codes and lots
s/n or lot number - Job Number lots 671494 - 672668

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2060-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2060-2017Class IITerminatedVoluntary: Firm initiated