Z-2613-2017Class II

ARCHITECT SHBG Reagent The ARCHITECT SHBG assay is a chemiluminescent microparticle immunoassay (CMIA) for...

Abbott Medical / initiated 2017-04-18 / Terminated / root cause: manufacturing process

Abbott Laboratories began a recall of ARCHITECT SHBG Reagent The ARCHITECT SHBG assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of sex hormone... on . FDA classified it as Class II and gives the reason as "Firm has identified that ARCHITECT SHBG results generated with the Potassium EDTA specimen tube type may demonstrate a negative shift relative to results generated with serum. There is a potential for falsely decreased...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2613-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
ARCHITECT SHBG Reagent The ARCHITECT SHBG assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of sex hormone binding globulin (SHBG) in human serum and plasma on the ARCHITECT iSystem. The ARCHITECT SHBG assay is used as an aid in the diagnosis of androgen disorders.
Product code
CDZ Radioimmunoassay, Testosterones And Dihydrotestosterone
Reason for recall
Firm has identified that ARCHITECT SHBG results generated with the Potassium EDTA specimen tube type may demonstrate a negative shift relative to results generated with serum. There is a potential for falsely decreased ARCHITECT SHBG results to be generated when using the ARCHITECT SHBG assay with the Potassium EDTA specimen tube type.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Abbott sent an Product Correction ( Immediate Action Required ) letter dated April 18, 2017, to all affected customers (with a response form) shipped via FedEx. Customers were instructed to discontinue use of the Potassium EDTA specimen tube type with the assay. All impacted US consignees were requested to return a customer reply form acknowledging that all the necessary actions, letter, were understood and completed. All impacted distributors were required to provide signed attestation ensuring that all their affected customers were sent the product correction letter informing them of this issue. Remaining inventory, at the firm's distribution centers, will have labeling corrected by inserting a kit stuffer informing customers that Potassium EDTA specimen tube type should not be used and providing customers will the recommend specimen types. If you or any of the health care providers you serve have any questions regarding this information, U.S. Customers please contact Customer Service at 1(877) - 4ABBOTT (available 24 hours a day, 7 days a week). Customers outside the U.S., please contact your local area Customer Service.
Quantity in commerce
7981 kits
Distribution
Worldwide Distribution - USA (nationwide) and to the countries of : ARGENTINA, ARMENIA, AUSTRIA, AZERBAIJAN, BANGLADESH, BARBADOS, BELARUS, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BRUNEI, CANADA, CHINA, CHILE, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DEM. REP. OF THE CON, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, GREECE, HUNGARY, INDIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JORDAN, KAZAKHSTAN, KUWAIT, LATVIA, LEBANON, LITHUANIA, LUXEMBOURG, MALAYSIA, MEXICO, MONTENEGRO, NAMIBIA, NETHERLANDS, NORWAY, OMAN, PAKISTAN, PARAGUAY, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UAE, UNITED KINGDOM, URUGUAY, VIETNAM
Product codes and lots
List number: 8K26-20, Lot Numbers: 01916E000, 00316F000, 01816G000, 00916I000, 07316I000, 00916L000 ;   List Number: 8K26-25, Lot Numbers: 01816E000, 00216F000, 01716G000, 00816I000, 07216I000

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2613-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2613-2017Class IITerminatedVoluntary: Firm initiated