Z-2671-2017Class II
Alaris PC Unit, Model 8015
CareFusion 303, Inc / initiated 2017-06-12 / Terminated / root cause: software
CareFusion 303, Inc. began a recall of Alaris PC Unit, Model 8015 on . FDA classified it as Class II and gives the reason as "BD initiated the recall of Alaris PC unit model 8015 after the firm identified five scenarios which can result in the occurrence of Systems Error Code 255-16-275 and can potentially result in interrupted infusions". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2671-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CareFusion 303, Inc
- Product
- Alaris PC Unit, Model 8015
- Product code
- FRN Pump, Infusion
- Reason for recall
- BD initiated the recall of Alaris PC unit model 8015 after the firm identified five scenarios which can result in the occurrence of Systems Error Code 255-16-275 and can potentially result in interrupted infusions.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- BD sent an Medical Device Safety Notification Letter dated June 12, 2017 , to customers to inform them that BD is issuing this Safety Notification letter to inform you of reports of customers experiencing a System Error (error code: 255-16-275) with Alaris System PC unit model 8015 that can result in interruption of infusions. Customers are informed of the issue, potential risk, and required actions to be taken by users and by BD. Customers with questions with the safety notification are instructed to call (888) 562-6018 or email [email protected]. Customers with adverse event reports are instructed to call (888) 812-3266 or email [email protected]. Customers with technical questions are instructed to call (888) 812-3229 or email [email protected]. For further questions, please call (858) 617-5925.
- Quantity in commerce
- 568,283 units
- Distribution
- Worldwide Distribution - USA (nationwide) and to the countries of : Europe, Australia, New Zealand, South Africa, Greater Asia, Middle East, and Canada.
- Product codes and lots
- Notification will be distributed to all Alaris PC Unit model 8015 customers with software versions greater than 9.12.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2671-2017 | Class II | Terminated | Voluntary: Firm initiated |