Pointe Scientific Chemistry Controls Level I and II LEVl: Amber glass vial w/PP white plastic cap. LEV2...
Medtest Holdings, Inc. began a recall of Pointe Scientific Chemistry Controls Level I and II LEVl: Amber glass vial w/PP white plastic cap. LEV2: Amber glass vial w/PP black plastic cap on . FDA classified it as Class III and gives the reason as "The original mean assigned to the control too high, upon reassignment with additional data the new assigned mean was outside of the originally published range". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2121-2017
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Horiba Instruments Incorporated
- Product
- Pointe Scientific Chemistry Controls Level I and II LEVl: Amber glass vial w/PP white plastic cap. LEV2: Amber glass vial w/PP black plastic cap.
- Product code
- JJY Multi-Analyte Controls, All Kinds (Assayed)
- Reason for recall
- The original mean assigned to the control too high, upon reassignment with additional data the new assigned mean was outside of the originally published range.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Customers were notified on 05/5/14by sending a technical bulletin (via email, fax, US Mail and posting on website) to all customers and distributors that had received lot 303104 of C7590, lot 303105 of C7591 or C7592 kits with either lot since product was approved to ship. All product shipped after 05/15/2014 included the updated package inserts. Any questions concerning the implementation of the new controls ranges may be directed to Pointe Scientific's Technical Service Department at 1-800-757-5313. Thank you for your continued support of Pointe Scientific.
- Quantity in commerce
- 132919 Vol vials
- Distribution
- Nationwide
- Product codes and lots
- Model Number(s): C7590-50 (Chemistry Control Level I), C7591-50 (Chemistry Control Level II), and C7592-100 (Chemistry Control Level I and II).
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2121-2017 | Class III | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY