Z-2121-2017Class III

Pointe Scientific Chemistry Controls Level I and II LEVl: Amber glass vial w/PP white plastic cap. LEV2...

Horiba Instruments Incorporated / initiated 2014-05-15 / Terminated / root cause: design

Medtest Holdings, Inc. began a recall of Pointe Scientific Chemistry Controls Level I and II LEVl: Amber glass vial w/PP white plastic cap. LEV2: Amber glass vial w/PP black plastic cap on . FDA classified it as Class III and gives the reason as "The original mean assigned to the control too high, upon reassignment with additional data the new assigned mean was outside of the originally published range". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2121-2017
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Horiba Instruments Incorporated
Product
Pointe Scientific Chemistry Controls Level I and II LEVl: Amber glass vial w/PP white plastic cap. LEV2: Amber glass vial w/PP black plastic cap.
Product code
JJY Multi-Analyte Controls, All Kinds (Assayed)
Reason for recall
The original mean assigned to the control too high, upon reassignment with additional data the new assigned mean was outside of the originally published range.
Root cause tag
design
FDA root cause
Component design/selection
Action
Customers were notified on 05/5/14by sending a technical bulletin (via email, fax, US Mail and posting on website) to all customers and distributors that had received lot 303104 of C7590, lot 303105 of C7591 or C7592 kits with either lot since product was approved to ship. All product shipped after 05/15/2014 included the updated package inserts. Any questions concerning the implementation of the new controls ranges may be directed to Pointe Scientific's Technical Service Department at 1-800-757-5313. Thank you for your continued support of Pointe Scientific.
Quantity in commerce
132919 Vol vials
Distribution
Nationwide
Product codes and lots
Model Number(s): C7590-50 (Chemistry Control Level I), C7591-50 (Chemistry Control Level II), and C7592-100 (Chemistry Control Level I and II).

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2121-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2121-2017Class IIITerminatedVoluntary: Firm initiated