Z-2707-2017Class II

The Schiller PB 1000, also known as Merge Hemo Patient Data Module (PDM), a major component of the Merge Hemo...

Merge Healthcare Incorporated / initiated 2016-03-10 / Terminated / root cause: software

Merge Healthcare, Inc. began a recall of The Schiller PB 1000, also known as Merge Hemo Patient Data Module (PDM), a major component of the Merge Hemo software on . FDA classified it as Class II and gives the reason as "If the PDM is utilized to output a pressure for computing fractional flow reserve (FFR), the analog output may not reflect the true representation of the selected pressure waveform being provided by the PDM to the Link...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2707-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merge Healthcare Incorporated
Product
The Schiller PB 1000, also known as Merge Hemo Patient Data Module (PDM), a major component of the Merge Hemo software.
Product code
DQK Computer, Diagnostic, Programmable
Reason for recall
If the PDM is utilized to output a pressure for computing fractional flow reserve (FFR), the analog output may not reflect the true representation of the selected pressure waveform being provided by the PDM to the Link Assembly which provides the analog output.
Root cause tag
software
FDA root cause
Software design
Action
The recalling firm issued recall letters dated 3/10/2016, 5/3/2016, and 8/5/2016 via email or via certified mail if the email address is not available.
Quantity in commerce
555 PDM's
Distribution
Distribution US nationwide.
Product codes and lots
All Schiller PB-1000 PDM's distributed on or before 8/11/2016.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2707-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2707-2017Class IITerminatedVoluntary: Firm initiated