Z-2180-2017Class II

Dolphin Inflation Device, REF 0185NA. Rx only, Sterilized using Ethylene Oxide. DOLPHIN is a single-use...

Vascular Solutions LLC / initiated 2017-04-12 / Terminated / root cause: packaging

Vascular Solutions, Inc. began a recall of Dolphin Inflation Device, REF 0185NA. Rx only, Sterilized using Ethylene Oxide. DOLPHIN is a single-use, sterile inflation device used in cardiovascular... on . FDA classified it as Class II and gives the reason as "The manufacturer of the DOLPHIN inflation device, Perouse Medical, has initiated a recall of the device due to a potential defect in the sterile barrier. Use of affected product could pose a risk of microbial...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2180-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Vascular Solutions LLC
Product
Dolphin Inflation Device, REF 0185NA. Rx only, Sterilized using Ethylene Oxide. DOLPHIN is a single-use, sterile inflation device used in cardiovascular procedures to pressurize (inflate) and depressurize (deflate) balloon catheters. The device itself consists of two components - the inflation device and large volume syringe (30 cc). The manually operation of the DOLPHIN is achieved by the manipulation of a large handle to drive a piston housed within the body of the device. Careful and controlled inflation is achieved by rotating the handle clockwise. During inflation a unique cam locking mechanism maintains pressure even if the user lets go of the device. Instantaneous deflation, regardless of balloon size, is made possible by the release of the dual locks located on device as well as a large, 30 cc syringe that is sold with each Dolphin Inflation Device. All the while during inflation or deflation pressure is displayed and can be monitored on a large analog gauge mounted on top of the device.
Product code
MAV Syringe, Balloon Inflation
Reason for recall
The manufacturer of the DOLPHIN inflation device, Perouse Medical, has initiated a recall of the device due to a potential defect in the sterile barrier. Use of affected product could pose a risk of microbial contamination, leading to infection.
Root cause tag
packaging
FDA root cause
Packaging
Action
Customers were sent a Vascular Solutions "Urgent Medical Device Removal letter" dated April 12, 2017. The letter described the problem and the product involved in the recall. Advised consignees to remove and secure all affected product and to complete and return the "VSI Account Inventory Form". For questions contact your local Vascular Solutions Sales Representative. For further questions, please call (763) 656-4300.
Quantity in commerce
4,110
Distribution
US Distribution including PR and to the states of : LA, TN, CA, SC, OR, WA, OK, NV, HI, OH, TX.
Product codes and lots
Lot Numbers: 15075668, 15075924, 15115426, 15125175, 16025324, 16035493, 16055261.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2180-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2180-2017Class IITerminatedVoluntary: Firm initiated