Z-2099-2017Class I

Zenith Alpha Thoracic Endovascular Graft The Zenith Alpha" Thoracic Endovascular Graft is indicated for the...

Cook Medical LLC / initiated 2017-03-22 / Terminated / root cause: design

Cook Medical Incorporated began a recall of Zenith Alpha Thoracic Endovascular Graft The Zenith Alpha" Thoracic Endovascular Graft is indicated for the endovascular treatment of patients with isolated... on . FDA classified it as Class I and gives the reason as "There have been five reports of graft thrombosis/occlusion during global commercial use, each following treatment for BTAI. One case resulted in patient death, and three cases resulted in reintervention". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2099-2017
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Medical LLC
Product
Zenith Alpha Thoracic Endovascular Graft The Zenith Alpha" Thoracic Endovascular Graft is indicated for the endovascular treatment of patients with isolated lesions of the descending thoracic aorta (not including dissections) having vascular anatomy suitable for endovascular repair including: " Iliac/femoral anatomy that is suitable for access with the required introduction systems, " Nonaneurysmal aortic segments (fixation sites) proximal and distal to the thoracic lesion: " with a length of at least 20 mm, and " with a diameter measured outer wall to outer wall of no greater than 42 mm and no less than 15 mm.
Product code
MIH System, Endovascular Graft, Aortic Aneurysm Treatment
Reason for recall
There have been five reports of graft thrombosis/occlusion during global commercial use, each following treatment for BTAI. One case resulted in patient death, and three cases resulted in reintervention.
Root cause tag
design
FDA root cause
Device Design
Action
The firm, Cook Medical, sent an "URGENT: MEDICAL DEVICE CORRECTION" letter dated March 22, 2017 to the affected consignees via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. On 6/22/2017, updated "URGENT: MEDICAL DEVICE CORRECTION AND REMOVAL" notifications were sent to the affected consignees via courier. The update included removal of the indication for use in BTAI, as follows: The Zenith Alpha Thoracic Endovascular Graft is indicated for the endovascular treatment of patients with aneurysms or ulcers of the descending thoracic aorta having vascular morphology suitable for endovascular repair including: " Iliac/femoral anatomy that is suitable for access with the required introduction systems " Nonaneurysmal aortic segments (fixation sites) proximal and distal to the aneurysm or ulcer: o with a length of at least 20 mm, and o with a diameter measured outer-wall-to-outer-wall of no greater than 42 mm and no less than 20 mm. Additionally, the following warning has been added to describe the thrombus risk that has been observed when the device is used to treat BTAI: " Risk of in-graft thrombus has been observed when the Zenith Alpha Thoracic Endovascular Graft has been used to treat BTAI. Because of the IFU correction to remove BTAI from the indication, it is necessary to remove specific sizes of this device (grafts with a proximal or distal diameter 18 22 mm) that would likely be used only for BTAI. This notice must be shared with appropriate personnel, including down to the user level, within your organization or to any organization where the potentially affected devices have been transferred. Should you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.
Quantity in commerce
4916
Distribution
Nationwide Distribution.
Product codes and lots
All lots. Catalog Numbers: ZTA-D-28-160-W ZTA-D-28-229-W ZTA-D-30-160-W ZTA-D-30-229-W ZTA-D-32-160-W ZTA-D-32-229-W ZTA-D-34-142-W ZTA-D-34-190-W ZTA-D-36-142-W ZTA-D-36-190-W ZTA-D-38-147-W ZTA-D-38-197-W ZTA-D-40-147-W ZTA-D-40-197-W ZTA-D-42-152-W ZTA-D-42-204-W ZTA-D-44-157-W ZTA-D-44-211-W ZTA-D-46-157-W ZTA-D-46-211-W ZTA-DE-18-104-W ZTA-DE-18-148-W ZTA-DE-20-104-W ZTA-DE-20-148-W ZTA-DE-22-104-W ZTA-DE-22-148-W ZTA-DE-24-104-W ZTA-DE-24-148-W ZTA-DE-26-104-W ZTA-DE-26-148-W ZTA-DE-28-108-W ZTA-DE-28-154-W ZTA-DE-30-108-W ZTA-DE-30-154-W ZTA-DE-32-108-W ZTA-DE-32-154-W ZTA-DE-34-112-W ZTA-DE-34-160-W ZTA-DE-36-112-W ZTA-DE-36-160-W ZTA-DE-38-141-W ZTA-DE-38-91-W ZTA-DE-40-141-W ZTA-DE-40-91-W ZTA-DE-42-146-W ZTA-DE-42-94-W ZTA-DE-44-151-W ZTA-DE-44-97-W ZTA-DE-46-151-W ZTA-DE-46-97-W ZTA-P-18-105-W ZTA-P-18-127-W ZTA-P-20-105-W ZTA-P-20-127-W ZTA-P-22-105-W ZTA-P-22-127-W ZTA-P-24-105-W ZTA-P-24-127-W ZTA-P-26-105-W ZTA-P-26-149-W ZTA-P-28-109-W ZTA-P-28-132-W ZTA-P-28-155-W ZTA-P-28-201-W ZTA-P-30-109-W ZTA-P-30-132-W ZTA-P-30-155-W ZTA-P-30-201-W ZTA-P-32-109-W ZTA-P-32-132-W ZTA-P-32-155-W ZTA-P-32-201-W ZTA-P-34-113-W ZTA-P-34-137-W ZTA-P-34-161-W ZTA-P-34-209-W ZTA-P-36-113-W ZTA-P-36-137-W ZTA-P-36-161-W ZTA-P-36-209-W ZTA-P-38-117-W ZTA-P-38-142-W ZTA-P-38-167-W ZTA-P-38-217-W ZTA-P-40-117-W ZTA-P-40-142-W ZTA-P-40-167-W ZTA-P-40-217-W ZTA-P-42-121-W ZTA-P-42-147-W ZTA-P-42-173-W ZTA-P-42-225-W ZTA-P-44-125-W ZTA-P-44-152-W ZTA-P-44-179-W ZTA-P-44-233-W ZTA-P-46-125-W ZTA-P-46-152-W ZTA-P-46-179-W ZTA-P-46-233-W ZTA-PT-22-18-105-W ZTA-PT-26-22-105-W ZTA-PT-30-26-108-W ZTA-PT-32-28-178-W ZTA-PT-32-28-201-W ZTA-PT-34-30-161-W ZTA-PT-34-30-209-W ZTA-PT-36-32-161-W ZTA-PT-36-32-209-W ZTA-PT-38-34-167-W ZTA-PT-38-34-217-W ZTA-PT-40-36-167-W ZTA-PT-40-36-217-W ZTA-PT-42-38-173-W ZTA-PT-42-38-225-W ZTA-PT-44-40-179-W ZTA-PT-44-40-233-W ZTA-PT-46-42-179-W ZTA-PT-46-42-233-W

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2099-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2099-2017Class ITerminatedVoluntary: Firm initiated