Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2055-2017Class II

Linear Straight Broach Handle (Rasp Handle) (Greatbatch Part # 510397; DJO Surgical Inc., # 803-03-159) and...

Greatbatch Medical / initiated 2017-04-23 / Terminated / root cause: manufacturing process

Greatbatch Medical began a recall of Linear Straight Broach Handle (Rasp Handle) (Greatbatch Part # 510397; DJO Surgical Inc., # 803-03-159) and EZ Clean" Linear Broach Handle Greatbatch Part #... on . FDA classified it as Class II and gives the reason as "Greatbatch Medical has voluntarily initiated a Recall for the Linear Straight Broach Handle (Rasp Handle) and EZ Clean" Linear Broach Handle. Greatbatch Medical has received reports that that the product is breaking at...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2055-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Greatbatch Medical
Product
Linear Straight Broach Handle (Rasp Handle) (Greatbatch Part # 510397; DJO Surgical Inc., # 803-03-159) and EZ Clean" Linear Broach Handle Greatbatch Part # T11824, DJO Surgical Inc. # 803-03-059).
Product code
HTQ Broach
Reason for recall
Greatbatch Medical has voluntarily initiated a Recall for the Linear Straight Broach Handle (Rasp Handle) and EZ Clean" Linear Broach Handle. Greatbatch Medical has received reports that that the product is breaking at the welds and that the locating pin at the distal end of the instrument is breaking off during use.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Greatbatch sent an Urgent Medical Device Recall letter dated April 23, 2017. The letter described the product being recalled and described Reason For Recall, Risk To Health, Replacement Devices, & Instructions to DJO Surgical Inc. Advised consignee to quarantine and return the affected product. If the product was further distributed they should contact the end users and share the recall information with them. Requested consignees to complete and return the "Field Recall Response Form" to Stericycle at [email protected] or fax it to 877-479-8063. For questions contact your Greatbatch Medical representative, or contact us at 1-763-951-8235, or e-mail us at [email protected]. For further questions, please call (763) 951-8184.
Quantity in commerce
205
Distribution
US Distribution to the state of : TX
Product codes and lots
Lot numbers : Linear Straight Broach Handle (Rasp Handle): 2988946, 3138186, 3185601, 3287315, 3300879, 3333172, 3433392, 3332697. EZ Clean" Linear Broach Handle: 3508491.