Absorb Bioresorbable Vascular Scaffold (BVS) system (Australia). The Absorb BVS is a temporary scaffold...
Abbott Vascular began a recall of Absorb Bioresorbable Vascular Scaffold (BVS) system (Australia). The Absorb BVS is a temporary scaffold indicated for improving coronary luminal diameter that... on . FDA classified it as Class I and gives the reason as "Abbott Vascular is initiating a recall of the Absorb Bioresorbable Vascular Scaffold (BVS) System due to studies showing elevated rates of major adverse events, specifically, myocardial infraction and scaffold...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2151-2017
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Vascular
- Product
- Absorb Bioresorbable Vascular Scaffold (BVS) system (Australia). The Absorb BVS is a temporary scaffold indicated for improving coronary luminal diameter that will eventually resorb and potentially facilitate normalization of vessel function in patients with ischemic heart disease due to de novo native coronary artery lesions. The treated lesion length should be less than the nominal scaffolding length (8 mm, 12 mm, 18 mm, 23 mm, 28 mm) with reference vessel diameters e 2.0 mm and d 3.8 mm.
- Product code
- PNY Absorbable Coronary Drug-Eluting Stent
- Reason for recall
- Abbott Vascular is initiating a recall of the Absorb Bioresorbable Vascular Scaffold (BVS) System due to studies showing elevated rates of major adverse events, specifically, myocardial infraction and scaffold thrombosis when compared to patients treated with the Xience metallic drug eluting stent.
- FDA root cause
- Other
- Action
- Abbott Vascular sent an Urgent Medical Device Recall and Hazard Alert letter dated April 26, 2017, to affected customers to inform them that the action is in response to recent concerns over data from some studies showing elevated rates of major adverse events, specifically, myocardial infraction and scaffold thrombosis when compared to patients treated with the Xience metallic drug eluting stent. Customers were informed of the actions to be taken. Customers were instructed to stop using these devices immediately, advise BVS patients to follow the recommendations for DAPT prescribed by their health care provider. Advise patients experiencing new cardiac symptoms to seek clinical care, review their inventory and complete the attached Facsimile Reply Form and display and share this notification with other relevant personnel within their organization. Customers with questions are instructed to contact their local Abbott Vascular Representative or Customer Service department.
- Quantity in commerce
- 590 units
- Distribution
- Australia
- Product codes and lots
- Australia: Part No. Description 1012462-08 2.5 x 08 mm Absorb BVS 1012462-12 2.5 x 12 mm Absorb BVS 1012462-18 2.5 x 18 mm Absorb BVS 1012462-23 2.5 x 23 mm Absorb BVS 1012462-28 2.5 x 28 mm Absorb BVS 1012463-08 3.0 x 08 mm Absorb BVS 1012463-12 3.0 x 12 mm Absorb BVS 1012463-18 3.0 x 18 mm Absorb BVS 1012463-23 3.0 x 23 mm Absorb BVS 1012463-28 3.0 x 28 mm Absorb BVS 1012464-12 3.5 x 12 mm Absorb BVS 1012464-18 3.5 x 18 mm Absorb BVS 1012464-23 3.5 x 23 mm Absorb BVS 1012464-28 3.5 x 28 mm Absorb BVS
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2151-2017 | Class I | Terminated | Voluntary: Firm initiated |