Z-2124-2017Class II

Microalbumin Reagent, Catalog No. 475100, Product Labeling A18520AL MA reagent, when used in conjunction with...

Siemens Healthcare Diagnostics Inc. / initiated 2017-04-24 / Terminated

Beckman Coulter Inc. began a recall of Microalbumin Reagent, Catalog No. 475100, Product Labeling A18520AL MA reagent, when used in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems... on . FDA classified it as Class II and gives the reason as "Beckman Coulter has confirmed that MA Reagent Lot M609458 may recover outside the lower limit of the published insert range for Beckman Coulter Urine Protein Control Level 1". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2124-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Microalbumin Reagent, Catalog No. 475100, Product Labeling A18520AL MA reagent, when used in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems MA Calibrator, is intended for quantitative determination of Albumin concentration in human urine
Product code
DCF Albumin, Antigen, Antiserum, Control
Reason for recall
Beckman Coulter has confirmed that MA Reagent Lot M609458 may recover outside the lower limit of the published insert range for Beckman Coulter Urine Protein Control Level 1.
FDA root cause
Under Investigation by firm
Action
Beckman Coulter sent an Urgent Medical Device Recall letter dated April 24, 2017, to affected customers to inform them that Beckman Coulter has confirmed that MA Reagent Lot M609458 may recover outside the lower limit of the published insert range for Beckman Coulter Urine Protein Control Level 1. Internal testing has confirmed that MA is recovering within the published insert range for Beckman Coulter Urine Protein Control Level 2. Customers are informed of the impact, actions to be taken, and resolution. Customers are instructed to complete and return the enclosed response form within 10 days. Customers with questions are instructed to Customer Support Center via http://www.beckmancoulter.com/customersupport/support, call 1-800-854-3633 in the United States and Canada, and outside US and Canada they are instructed to contact their local Beckman Coulter Representative.
Quantity in commerce
3,158 units total (1,957 units in US)
Distribution
Worldwide Distribution - US (nationwide) and Internationally to Australia, Belgium, Brunei, Bulgaria, Canada, China, Colombia, Czech Republic, France, Germany, Guatemala, India, Italy, Japan, Mayotte, Mexico, Netherlands, Oman, Philippines, Portugal, Qatar, Saudi Arabia, South Africa, Spain, Switzerland, Taiwan, and United Kingdom.
Product codes and lots
M609458

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2124-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2124-2017Class IITerminatedVoluntary: Firm initiated