510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2543-2017Class II

Sutures. ENDO STITCH: ENDO STITCH POLYSORB 0 48" U/D DLU SU, Item Number: 170056; ENDO STITCH POLYSORB 0 48"...

Medtronic, Inc. / initiated 2017-05-03 / Terminated / root cause: manufacturing process

Medtronic began a recall of Sutures. ENDO STITCH: ENDO STITCH POLYSORB 0 48" U/D DLU SU, Item Number: 170056; ENDO STITCH POLYSORB 0 48" VIO DLU SU, Item Number: 170052; ENDO STITCH... on . FDA classified it as Class II and gives the reason as "Potential for an incomplete seal on the inner package, which may result in premature degradation of the suture material". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2543-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Sutures. ENDO STITCH: ENDO STITCH POLYSORB 0 48" U/D DLU SU, Item Number: 170056; ENDO STITCH POLYSORB 0 48" VIO DLU SU, Item Number: 170052; ENDO STITCH POLYSORB 2/0 48" U/D DLU SU, Item Number: 170057; ENDO STITCH POLYSORB 2/0 48" VIO DLU SU, Item Number: 170053
Product code
GAM Suture, Absorbable, Synthetic, Polyglycolic Acid
Reason for recall
Potential for an incomplete seal on the inner package, which may result in premature degradation of the suture material.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
All consignees were notified via Federal Express or certified mail on May 3, 2017, and the letter informs customers of the potential for an incomplete seal on the inner package and the actions they should take. Customers are requested to acknowledge receipt and understanding of the Urgent Field Corrective Action Notice.
Quantity in commerce
478,465 (total for all products)
Distribution
Worldwide distribution. US Nationwide including PR, and countries of AU, BE, CA, CL, CN, IL, IT, JP, KR, MX, PA, SA, SG, and TR.
Product codes and lots
ENDO STITCH* POLYSORB* 0 48" U/D DLU SU J4M0853X J4M0855X J5A0192X ENDO STITCH* POLYSORB* 0 48" VIO DLU SU J4L1477X ENDO STITCH*POLYSORB* 2/0 48" U/D DLU SU J5A1097X ENDO STITCH*POLYSORB* 2/0 48" VIO DLU SU J5A0702X J5A0703X J5A0704X J5A0705X J5A1206X J5A1207X

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2543-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2541-2017Class IITerminatedVoluntary: Firm initiated
Z-2542-2017Class IITerminatedVoluntary: Firm initiated
Z-2543-2017Class IITerminatedVoluntary: Firm initiated
Z-2544-2017Class IITerminatedVoluntary: Firm initiated
Z-2545-2017Class IITerminatedVoluntary: Firm initiated
Z-2546-2017Class IITerminatedVoluntary: Firm initiated
Z-2547-2017Class IITerminatedVoluntary: Firm initiated
Z-2548-2017Class IITerminatedVoluntary: Firm initiated