CytoGuard Closed Luer Connector, Intravenous access 2 cartons of 50 units each (100 units per case)
B Braun Medical Inc began a recall of CytoGuard Closed Luer Connector, Intravenous access 2 cartons of 50 units each (100 units per case) on . FDA classified it as Class II and gives the reason as "B. Braun is voluntarily recalling 13 lots of CytoGuard Closed Luer Connector due to a potential for some blisters to be punctured resulting in a compromised sterility barrier". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2155-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- B Braun Medical Inc
- Product
- CytoGuard Closed Luer Connector, Intravenous access 2 cartons of 50 units each (100 units per case)
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- B. Braun is voluntarily recalling 13 lots of CytoGuard Closed Luer Connector due to a potential for some blisters to be punctured resulting in a compromised sterility barrier.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- B. Braun mailed an Urgent Medical Device Recall Notice to affected customers to inform them of the issue. Customers were asked to determine their current inventory and to not destroy any of the product; however, product should be discontinued immediately and quarantined and return the acknowledgement form for product return.
- Quantity in commerce
- 417,600 units
- Distribution
- United States Nationwide distribution
- Product codes and lots
- Lot # 1240156001, 1240156002, 1240156003, 1240156004, 1370249001, 1370251101, 1370257101, 1370259201, 1370268401, 1370270401, 1370273801, 1470311101, and 1570315301