UniCel DxH 600 Coulter Cellular Analysis System, Catalog No. B23858
Beckman Coulter Inc. began a recall of UniCel DxH 600 Coulter Cellular Analysis System, Catalog No. B23858 on . FDA classified it as Class II and gives the reason as "There is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The events associated...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2158-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- UniCel DxH 600 Coulter Cellular Analysis System, Catalog No. B23858.
- Product code
- GKZ Counter, Differential Cell
- Reason for recall
- There is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The events associated with the unusual light scatter pattern may be incorrectly removed from analysis which can result in an erroneous differential result.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- An Urgent Medical Device Recall letter dated 4/24/17 was sent to customers to inform them that there is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The events associated with the unusual light scatter pattern may be incorrectly removed from analysis which can result in an erroneous differential result. The letter informs the customers of the issue, impact, action, and resolution. Customers are instructed to complete and return the enclosed response form within 10 days. Customers with any questions regarding the notice are instructed to contact Customer Support Center at Http://www.beckmancoulter.com/customersupport/support, 800-526-7694 in the US and Canada, customers outside the U.S. and Canada are instructed to contact their local Beckman Coulter representative.
- Quantity in commerce
- 1408 units total (998 units in US)
- Distribution
- Worldwide distribution: US nationwide, Algeria, Andorra, Austria, Australia, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Canada, China, Colombia, Costa Rica, Cote d'lvoire, Croatia, Czech Republic, Ecuador, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan , Kuwait, Lebanon, Lithuania ,Macao, Malaysia, Mayotte, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania , Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia , South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Republic, United Kingdom, Uruguay and Vietnam.
- Product codes and lots
- all software versions
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2157-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2158-2017 | Class II | Terminated | Voluntary: Firm initiated |