Z-2167-2017Class III

Liquid Urine Control Level 3 For Professional Use in the quality control monitoring of In-Vitro Diagnostic...

Randox Laboratories, Limited / initiated 2017-04-27 / Terminated / root cause: labeling and UDI

Randox Laboratories began a recall of Liquid Urine Control Level 3 For Professional Use in the quality control monitoring of In-Vitro Diagnostic Urinary Clinical Assays on a range of clinical... on . FDA classified it as Class III and gives the reason as "According to Randox Laboratories an investigation indicates a labelling error for 2 of the test methods listed for quantification of Total Protein (urine). The values for the mean of all instruments are incorrect by a...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2167-2017
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Randox Laboratories, Limited
Product
Liquid Urine Control Level 3 For Professional Use in the quality control monitoring of In-Vitro Diagnostic Urinary Clinical Assays on a range of clinical chemistry systems.
Product code
JJY Multi-Analyte Controls, All Kinds (Assayed)
Reason for recall
According to Randox Laboratories an investigation indicates a labelling error for 2 of the test methods listed for quantification of Total Protein (urine). The values for the mean of all instruments are incorrect by a factor of 10. Instrument specific values are correct. This does not affect the performance or stability claims of the product. As control results will be outside of the quoted range this may result in a delay in reporting the sample test results. However as the correct instrument specific means are also included in the value sheet this delay is unlikely. Customers have been referred to their Medical Director for further advice
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
Randox Laboratories will contact Customers directly by e-mail and will be followed up with once every 2 weeks over a period of 6 weeks. There are no plans in place to follow up with customers after the 6 week period. For further questions, please call (304) 728-2890.
Quantity in commerce
198 kits
Distribution
Worldwide Distribution - US Distribution to Puerto Rico and to the countries of : Canada Kuwait Chile Poland China Romania France Saudi Arabia Guatemala Slovenia Iran UAE India UK Italy Uruguay Iraq
Product codes and lots
Catalogue number UC5075 Batch number 907UC

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2167-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2167-2017Class IIITerminatedVoluntary: Firm initiated