10-1398M - SpF-PLUS Implantable Stimulator. Indicated as a spinal fusion adjunct to increase the probability...
EBI Patient Care, Inc. began a recall of 10-1398M - SpF-PLUS Implantable Stimulator. Indicated as a spinal fusion adjunct to increase the probability of fusion success in 1 or 2 levels on . FDA classified it as Class I and gives the reason as "Zimmer Biomet is conducting a medical device field action for the SpF PLUS-Mini and SpF XL IIB Implantable Spinal Fusion Stimulator products due to higher than allowed cytotoxicity levels found during routine process...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2119-2017
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- EBI Patient Care, Inc.
- Product
- 10-1398M - SpF-PLUS Implantable Stimulator. Indicated as a spinal fusion adjunct to increase the probability of fusion success in 1 or 2 levels.
- Product code
- LOE Stimulator, Invasive Bone Growth
- Reason for recall
- Zimmer Biomet is conducting a medical device field action for the SpF PLUS-Mini and SpF XL IIB Implantable Spinal Fusion Stimulator products due to higher than allowed cytotoxicity levels found during routine process monitoring with the product.
- FDA root cause
- Other
- Action
- An Urgent Medical Device Recall letter, dated April 20, 2017, was sent to the firm's consignees. The letter identified the affected product and the reason for the recall, as well as the risks involved. Consignees were instructed to quarantine affected product in their inventory and complete the Certification of Acknowledgement portion and the Inventory Return Certification Form in the letter. Affected product must be returned immediately along with the completed Inventory Return Certification Form. Questions or concerns should be directed to the customer call center at 1-800-447-3625 or via email to [email protected].
- Quantity in commerce
- 33 units total
- Distribution
- Nationwide distribution only.
- Product codes and lots
- Serial Numbers: 410093 410094 410096 410103 410111 410115 410119 410148 410151 410158, and 410171.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2118-2017 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-2119-2017 | Class I | Terminated | Voluntary: Firm initiated |