Z-2162-2017Class II

CORFLO, PEG Adaptor Repair Kit with ENFit Connectors, Size: 16FR

Halyard Health / initiated 2017-02-03 / Terminated / root cause: design

CORPAK MedSystems, Inc. began a recall of CORFLO, PEG Adaptor Repair Kit with ENFit Connectors, Size: 16FR; Product Usage: CORFLO PEG Tubes are intended for delivery of enteral nutrition, water and... on . FDA classified it as Class II and gives the reason as "Firm is initiating voluntary correction of CORFLO PEG with ENFit. Barb connector (stem) of the adapter may break unexpectedly when exposed to stress". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2162-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Halyard Health
Product
CORFLO, PEG Adaptor Repair Kit with ENFit Connectors, Size: 16FR; Product Usage: CORFLO PEG Tubes are intended for delivery of enteral nutrition, water and medications into the stomach. The PEG adapter is the connection point between the tubing and the feeding set/syringe. The adapter is designed as a replaceable component.
Product code
KNT Tubes, Gastrointestinal (And Accessories)
Reason for recall
Firm is initiating voluntary correction of CORFLO PEG with ENFit. Barb connector (stem) of the adapter may break unexpectedly when exposed to stress.
Root cause tag
design
FDA root cause
Device Design
Action
Firm provided consignees with field correction on 2/2/2017 by letter mailed via FedEx. Within the notice, the firm provided additional instructions for handling the PEG Y-Adapter to help reduce the likelihood for exposing the stem to excess stress that could break the connector. Consignees were instructed to forward the field correction notice to all staff members or end-users who use the ENFit Y-Adapter.
Quantity in commerce
2 cases (5 units per case)
Distribution
Worldwide distribution in the states of NH, VT, CA and the countries of UK, Ireland, France, Austria, Norway, Sweden, New Zealand
Product codes and lots
Catalogue Numbers: 50-6116

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2162-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2161-2017Class IITerminatedVoluntary: Firm initiated
Z-2162-2017Class IITerminatedVoluntary: Firm initiated
Z-2163-2017Class IITerminatedVoluntary: Firm initiated