CORFLO PEG Kit, Pull 16FR
CORPAK MedSystems, Inc. began a recall of CORFLO PEG Kit, Pull 16FR. Product Usage: CORFLO PEG Tubes are intended for delivery of enteral nutrition, water and medications into the stomach. The PEG... on . FDA classified it as Class II and gives the reason as "Firm is initiating voluntary correction of CORFLO PEG with ENFit. Barb connector (stem) of the adapter may break unexpectedly when exposed to stress". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2163-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Halyard Health
- Product
- CORFLO PEG Kit, Pull 16FR. Product Usage: CORFLO PEG Tubes are intended for delivery of enteral nutrition, water and medications into the stomach. The PEG adapter is the connection point between the tubing and the feeding set/syringe. The adapter is designed as a replaceable component.
- Product code
- KNT Tubes, Gastrointestinal (And Accessories)
- Reason for recall
- Firm is initiating voluntary correction of CORFLO PEG with ENFit. Barb connector (stem) of the adapter may break unexpectedly when exposed to stress.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Firm provided consignees with field correction on 2/2/2017 by letter mailed via FedEx. Within the notice, the firm provided additional instructions for handling the PEG Y-Adapter to help reduce the likelihood for exposing the stem to excess stress that could break the connector. Consignees were instructed to forward the field correction notice to all staff members or end-users who use the ENFit Y-Adapter.
- Quantity in commerce
- 8 cases (2 units per case)
- Distribution
- Worldwide distribution in the states of NH, VT, CA and the countries of UK, Ireland, France, Austria, Norway, Sweden, New Zealand
- Product codes and lots
- Catalogue Numbers: 50-6016
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2161-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2162-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2163-2017 | Class II | Terminated | Voluntary: Firm initiated |