Cobas connection module (CCM) - OUTPUT UNIT The module CCM enables a unidirectional connection of the...
Roche Diagnostics Corporation began a recall of Cobas connection module (CCM) - OUTPUT UNIT The module CCM enables a unidirectional connection of the pre-analytical systems cobas p 612/ p 512 to analysis... on . FDA classified it as Class II and gives the reason as "a gap between the conveyor belt and the rack tray table (plastic part) may cause vibration of the 5 position racks during the transport. This issue can pose a potential risk to involved operators and can also lead to...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2223-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Cobas connection module (CCM) - OUTPUT UNIT The module CCM enables a unidirectional connection of the pre-analytical systems cobas p 612/ p 512 to analysis instruments from Roche Diagnostics, Hitachi or Sysmex by means of the Aloka transportation system
- Product code
- JQP Calculator/Data Processing Module, For Clinical Use
- Reason for recall
- a gap between the conveyor belt and the rack tray table (plastic part) may cause vibration of the 5 position racks during the transport. This issue can pose a potential risk to involved operators and can also lead to potential cross contamination of samples in the affected racks.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Roche sent an Urgent Medical Device Correction letter dated May 9, 2017, to all affected consignees via e-mail. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Consignees were asked to complete the attached form and fax it to 877-906-8982 or email to [email protected]. Consignees with questions were instructed to contact the Roche Support Network Customer Support Center at 1-800-428-2336.
- Quantity in commerce
- 3
- Distribution
- US distribution to WA only.
- Product codes and lots
- Serial Number 204C5690 204H0157 204H0158
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2223-2017 | Class II | Terminated | Voluntary: Firm initiated |