DePuy Mitek LATARJET EXPERIENCE Top Hat Tap
DePuy Mitek, Inc., a Johnson & Johnson Co. began a recall of DePuy Mitek LATARJET EXPERIENCE Top Hat Tap Product Code: 288203 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder... on . FDA classified it as Class II and gives the reason as "Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to screw". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2618-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- DePuy Mitek, Inc.
- Product
- DePuy Mitek LATARJET EXPERIENCE Top Hat Tap Product Code: 288203 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by supporting the anteroinferior glenoid with a bony graft
- Product code
- HWC Screw, Fixation, Bone
- Reason for recall
- Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to screw
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- DePuy Mitek issued recall letter dated May 12, 2017 advising of the problem and requesting return of the product. A Response Form is to be completed and returned. Questions: contact your DePuy Synthes Mitek Sales Consultant or Carolyn Somerville, DePuy Mitek, Inc. (via telephone: 508.828.3647 Fax: 508.828.3762 or via email: [email protected]).
- Quantity in commerce
- 153 units
- Distribution
- US: AZ, CA, CO, GA, MA, MI, OH, WA Foreign: Austria, Australia, Belgium, France, Germany, Spain, Switzerland, Netherlands, Poland, UK
- Product codes and lots
- GTN: (01)10886705026814 Lot codes: 16D01, 16D02, 16E01, 16J01, 16R01, 17A01, 17A02, 17C03
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2615-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2616-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2617-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2618-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2619-2017 | Class II | Terminated | Voluntary: Firm initiated |