Oxygen Masks, Elongated (under chin), Adult Med Conc., No Tubing, Product Code 104-E
ConvaTec, Inc began a recall of Oxygen Masks, Elongated (under chin), Adult Med Conc., No Tubing, Product Code 104-E on . FDA classified it as Class II and gives the reason as "Reports of tubing detaching from oxygen masks either prior to use (in packaging) or during use were made by different end users and for different batches. The failures were caused by an improper solvent bond between the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2228-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- ConvaTec Limited
- Product
- Oxygen Masks, Elongated (under chin), Adult Med Conc., No Tubing, Product Code 104-E;
- Product code
- BYG Mask, Oxygen
- Reason for recall
- Reports of tubing detaching from oxygen masks either prior to use (in packaging) or during use were made by different end users and for different batches. The failures were caused by an improper solvent bond between the oxygen tubing and the oxygen mask connector. While the defect affects less than 2% of recalled product, there is a potential for harm when oxygen supply to the user is disrupted.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Recall Letters were sent to consignees on May 10, 2017. The letter requested that they immediately stop distribution and quarantine all affected lots. The letter also requested a sub-recall if the product had been further distributed. Also attached to the letter is a Recall Response Form to be returned.
- Quantity in commerce
- 816,709
- Distribution
- CA, CT, FL, GA, IL, MO, NJ, NY, NC, OH, PA, RI, VA , Puerto Rico, Australia, Belgium, Canada, Columbia, Czech, Denmark, Finland, France, Germany, Greece, Israel, Italy, Japan, Jordan, Malaysia, Malta, Netherlands, Norway, Rwanda, Saudi Arabia, Sweden, Switzerland, United Kingdom, and Vietnam
- Product codes and lots
- Lot Number 102153, 102339, 102941, 103176, 103453, 103718, 103919, 104471, 105034, 105282, 105427, 105567, 105996, 106960, 108658, 112024, 112530
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2228-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2229-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2230-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2231-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2232-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2233-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2234-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2235-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2236-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2237-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2238-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2239-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2240-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2241-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2242-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2243-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2244-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2245-2017 | Class II | Terminated | Voluntary: Firm initiated |