Z-2700-2017Class II
Alaris Pump Module model 8100
CareFusion 303, Inc / initiated 2017-06-12 / Terminated / root cause: design
CareFusion 303, Inc. began a recall of Alaris Pump Module model 8100 on . FDA classified it as Class II and gives the reason as "There is a potential risk that could cause unintended flow in the older, centered sear door latch design in the Alaris Pump module model 8100". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2700-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CareFusion 303, Inc
- Product
- Alaris Pump Module model 8100
- Product code
- FRN Pump, Infusion
- Reason for recall
- There is a potential risk that could cause unintended flow in the older, centered sear door latch design in the Alaris Pump module model 8100.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- An Urgent Medical Device Recall letter will be sent on 6/12/17 to customers to inform them that BD is issuing this letter to inform you of a potential risk associated with the Alaris Pump module. Firm identified a specific scenario that could cause unintended flow in the older, centered sear door latch design in the Alaris Pump module model 8100. This scenario is reproducible under the following situations: a) the user does not close the roller clamp on the IV administration set before the pump door is opened as recommended, and b) opening the pump door using an atypical technique with the door latch with the centered sear design. The letter instructs customers If their device has a centered-sear that was manufactured between June 2002 to June 2004, clearly mark and sequester (e.g. Biomed department) the Alaris Pump module that exhibited the issue. Notify CareFusion Customer Advocacy at 888-812-3266 or [email protected] to coordinate an onsite remediation or send the devices to the BD Service Depot.
- Quantity in commerce
- 35,940 units
- Distribution
- US and Canada
- Product codes and lots
- Notification will be distributed to all Alaris Pumps customers with pumps manufactured between June 2002 through June 2004.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2700-2017 | Class II | Terminated | Voluntary: Firm initiated |