Z-2700-2017Class II

Alaris Pump Module model 8100

CareFusion 303, Inc / initiated 2017-06-12 / Terminated / root cause: design

CareFusion 303, Inc. began a recall of Alaris Pump Module model 8100 on . FDA classified it as Class II and gives the reason as "There is a potential risk that could cause unintended flow in the older, centered sear door latch design in the Alaris Pump module model 8100". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2700-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
CareFusion 303, Inc
Product
Alaris Pump Module model 8100
Product code
FRN Pump, Infusion
Reason for recall
There is a potential risk that could cause unintended flow in the older, centered sear door latch design in the Alaris Pump module model 8100.
Root cause tag
design
FDA root cause
Device Design
Action
An Urgent Medical Device Recall letter will be sent on 6/12/17 to customers to inform them that BD is issuing this letter to inform you of a potential risk associated with the Alaris Pump module. Firm identified a specific scenario that could cause unintended flow in the older, centered sear door latch design in the Alaris Pump module model 8100. This scenario is reproducible under the following situations: a) the user does not close the roller clamp on the IV administration set before the pump door is opened as recommended, and b) opening the pump door using an atypical technique with the door latch with the centered sear design. The letter instructs customers If their device has a centered-sear that was manufactured between June 2002 to June 2004, clearly mark and sequester (e.g. Biomed department) the Alaris Pump module that exhibited the issue. Notify CareFusion Customer Advocacy at 888-812-3266 or [email protected] to coordinate an onsite remediation or send the devices to the BD Service Depot.
Quantity in commerce
35,940 units
Distribution
US and Canada
Product codes and lots
Notification will be distributed to all Alaris Pumps customers with pumps manufactured between June 2002 through June 2004.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2700-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2700-2017Class IITerminatedVoluntary: Firm initiated