GE Healthcare, LLC began a recall of Infant Warmer System (IWS) on . FDA classified it as Class II and gives the reason as "GE Healthcare has recently become aware of a potential safety issue related to loose screws in the Heater Head of the Infant Warmer System (IWS). Hot screws from the Heater Head of the IWS could fall onto the bed if the...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2323-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Infant Warmer System (IWS)
- Product code
- FMT Warmer, Infant Radiant
- Reason for recall
- GE Healthcare has recently become aware of a potential safety issue related to loose screws in the Heater Head of the Infant Warmer System (IWS). Hot screws from the Heater Head of the IWS could fall onto the bed if the Heater Head assembly has been improperly serviced. This situation can be clinically hazardous because thermal injury to a patient could result. Two injuries have been reported as a result of this issue.
- FDA root cause
- Under Investigation by firm
- Action
- GE Healthcare sent an Urgent Medical Device Correction letter dated May 5, 2017, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Product Correction : Attached to this letter, we provide instructions as part of a Service Manual Addendum on how to correct the issue. Please add this new Addendum to the Service Manual of your device(s) and train the affected users accordingly. Please acknowledge that you have received this letter and that you understand that an action needs to be taken on your part to correct this issue by filling out and returning the attached Customer Response form. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative. Please be assured that maintaining a high level of safety and quality is our highest priority. If you have any questions, please contact us immediately per the contact information above.
- Quantity in commerce
- 34986
- Distribution
- Worldwide Distribution - USA (nationwide) Distribution.
- Product codes and lots
- Medical device listing number: D103053, D103054, D103056, D103057, D183529
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2323-2017 | Class II | Terminated | Voluntary: Firm initiated |