MesaLabs Conductivity Standard Solution, 14 mS/cm, 32 ounces (950 ml). Product No. : 02.0027, 020510 (carton...
Mesa Laboratories, Inc. began a recall of MesaLabs Conductivity Standard Solution, 14 mS/cm, 32 ounces (950 ml). Product No. : 02.0027, 020510 (carton of 6 of 02.0027). Polyethylene bottle... on . FDA classified it as Class II and gives the reason as "Mesa Laboratories is recalling Conductivity/TDS Calibrator Solution due to failure to meet product specifications". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2159-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Mesa Laboratories Inc
- Product
- MesaLabs Conductivity Standard Solution, 14 mS/cm, 32 ounces (950 ml). Product No. : 02.0027, 020510 (carton of 6 of 02.0027). Polyethylene bottle Conductivity/TDS Calibrator solutions are a secondary standard solution used for the calibration of conductivity/TDS cells together with conductivity/TDS measurement instruments. The conductivity/TDS cells and instruments may be indicated for calibrating the conductivity measurement function of hemodialysis machines and water purification equipment for hemodialysis, or for verifying proper function of hemodialysis machine or water purification equipment measurement functions. These solutions are used remotely from the hemodialysis machine or water purification equipment, and do not come into contact with the patient.
- Product code
- FKH Solution-Test Standard-Conductivity, Dialysis
- Reason for recall
- Mesa Laboratories is recalling Conductivity/TDS Calibrator Solution due to failure to meet product specifications.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- MesaLabs sent an Device Correction Notice dated May 12, 2017, via certified letter to destroy product by pouring it down the drain and free replacements will be shipped. For further questions, please call (303) 987-8000 ext. 120.
- Quantity in commerce
- 9900
- Distribution
- USA (nationwide) Distribution and Internationally to one customer in Canada, and govt/VA/military
- Product codes and lots
- Lot number: ML-C14-1352
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2159-2017 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY