Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2159-2017Class II

MesaLabs Conductivity Standard Solution, 14 mS/cm, 32 ounces (950 ml). Product No. : 02.0027, 020510 (carton...

Mesa Laboratories Inc / initiated 2017-05-12 / Terminated / root cause: manufacturing process

Mesa Laboratories, Inc. began a recall of MesaLabs Conductivity Standard Solution, 14 mS/cm, 32 ounces (950 ml). Product No. : 02.0027, 020510 (carton of 6 of 02.0027). Polyethylene bottle... on . FDA classified it as Class II and gives the reason as "Mesa Laboratories is recalling Conductivity/TDS Calibrator Solution due to failure to meet product specifications". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2159-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Mesa Laboratories Inc
Product
MesaLabs Conductivity Standard Solution, 14 mS/cm, 32 ounces (950 ml). Product No. : 02.0027, 020510 (carton of 6 of 02.0027). Polyethylene bottle Conductivity/TDS Calibrator solutions are a secondary standard solution used for the calibration of conductivity/TDS cells together with conductivity/TDS measurement instruments. The conductivity/TDS cells and instruments may be indicated for calibrating the conductivity measurement function of hemodialysis machines and water purification equipment for hemodialysis, or for verifying proper function of hemodialysis machine or water purification equipment measurement functions. These solutions are used remotely from the hemodialysis machine or water purification equipment, and do not come into contact with the patient.
Product code
FKH Solution-Test Standard-Conductivity, Dialysis
Reason for recall
Mesa Laboratories is recalling Conductivity/TDS Calibrator Solution due to failure to meet product specifications.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
MesaLabs sent an Device Correction Notice dated May 12, 2017, via certified letter to destroy product by pouring it down the drain and free replacements will be shipped. For further questions, please call (303) 987-8000 ext. 120.
Quantity in commerce
9900
Distribution
USA (nationwide) Distribution and Internationally to one customer in Canada, and govt/VA/military
Product codes and lots
Lot number: ML-C14-1352

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2159-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2159-2017Class IITerminatedVoluntary: Firm initiated