Z-2622-2017Class II
Covidien Endo Peanut 5mm Auto Suture Blunt Dissector Item Code: 173019 The instrument has indications in...
Medtronic, Inc. / initiated 2017-05-09 / Terminated / root cause: manufacturing process
Covidien LLC began a recall of Covidien Endo Peanut 5mm Auto Suture Blunt Dissector Item Code: 173019 The instrument has indications in endoscopic surgery for swabbing small amounts of fluid... on . FDA classified it as Class II and gives the reason as "Cotton tip of the device may disengage due to insufficient adhesive". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2622-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Covidien Endo Peanut 5mm Auto Suture Blunt Dissector Item Code: 173019 The instrument has indications in endoscopic surgery for swabbing small amounts of fluid or for blunt dissection of soft tissue(s) and structures, or as an aid in controlling minor intraoperative bleeding
- Product code
- GCJ Laparoscope, General & Plastic Surgery
- Reason for recall
- Cotton tip of the device may disengage due to insufficient adhesive
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Medtronic consignees were notified via Federal Express or certified mail on May 9, 2017, and the letter informs customers of the potential for the tip of the device to disengage from the device and the actions they should take. Accounts requested to discontinue use, quarantine and return product. Customers are requested to complete and acknowledge receipt of the Urgent Field Corrective Action Notice. Questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2. For questions regarding this recall call 203-492-5000.
- Quantity in commerce
- 194,851
- Distribution
- Nationwide Distribution
- Product codes and lots
- Lot Numbers: M5K04X to M7A01X Exp Date: 2020-October through 2022-January
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2622-2017 | Class II | Terminated | Voluntary: Firm initiated |