Large Stealth Chamber, Catalog Numbers: DS50-001-XL-T, DS50-002-XL-T, DS50-003-XL-T, DS50-102-XL-T...
Iba Dosimetry began a recall of Large Stealth Chamber, Catalog Numbers: DS50-001-XL-T, DS50-002-XL-T, DS50-003-XL-T, DS50-102-XL-T, DS50-103-XL-T on . FDA classified it as Class II and gives the reason as "When the Stealth Chamber is used as a reference detector in relative dosimetry, deviation in the measured PDDs at shallow depths and geometric shadowing of measured profiles may appear". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2220-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- IBA Dosimetry GmbH
- Product
- Large Stealth Chamber, Catalog Numbers: DS50-001-XL-T, DS50-002-XL-T, DS50-003-XL-T, DS50-102-XL-T, DS50-103-XL-T
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- When the Stealth Chamber is used as a reference detector in relative dosimetry, deviation in the measured PDDs at shallow depths and geometric shadowing of measured profiles may appear.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- IBA Dosimetry sent an Field Safety Notice dated April 21, 2017. The letter informed the consignee of the issue and provided the following instruction: If a Stealth Chamber was used during commissioning of Linear Accelerator and Treatment Planning System, the consignee was asked to consider referenced deviations for larger field sizes (>20 x 20 cm) and higher energies (>60MV). They were also instructed to recheck the influence of the PDD deviation and insufficient clearance for 40 x 40 profile on their beam model and its clinical relevancy. For further questions please call (901) 268-6430. Contact information Should you have questions or require additional information, please contact the Customer Service Team at:: South and Middle America, USA, and Canada : IBA Dosimetry America Bartlett, TN, USA US [email protected] .com Phone: +1 901 386-2242 Fax: +1 901 382-9453 Asia Pacific, A Australia and New Zealand: IBA China, Dosimetry Dept. Beijing, China Service [email protected] .com Phone: +86 10 8080 9107 Fax: + 86 10 80809298 All other countries: IBA Dosimetry GmbH, Service Dept. Schwarzenbruck , Germany [email protected] .com Phone: +49 9128 607-38 Fax: +49 9128 607-26
- Quantity in commerce
- 110 units
- Distribution
- Worldwide Distribution - USA (nationwide) and to the countries of : Argentina, Australia, Austria, Canada, Cyprus, France, Germany, Hong Kong, Ireland, Islamic Republic of, Israel, Italy, Japan, Kingdom of Saudi Arabia, Kuwait, New Zealand, People's Republic of China, Philippines, Russia, Slovakia, South Africa, South Korea, Spain, Sweden, Taiwan, R.O.C., and United Kingdom.
- Product codes and lots
- DS50-001-XL-T, Serial Numbers: 24887, 25084, 25085 DS50-002-XL-T, Serial Numbers: 24185, 25219, 25470, 24073, 24187, 24480, 24888, 25089, 24889, 24184, 24813, 24814, 24866, 24867, 24890, 25088, 25220, 24482, 24186, 25090, 24626, 24074, 24189, 25218, 25468, 25469 DS50-003-XL-T, Serial Numbers: 24437, 24260, 24639, 25472, 24257, 24291, 24630, 24640, 24868, 25093, 25478, 24180, 24182, 24259, 24323, 24439, 25222, 24179, 24290, 24293, 24633, 24637, 25091, 25223, 24627, 24077, 24634, 24438, 25476, 25221, 24292, 24263, 24322, 24435, 24628, 24642, 25094, 25479, 24177, 24178, 24641, 25481, 24261, 24076, 24871, 25224, 24078, 24079, 24080, 24175, 24176, 24183, 24256, 24258, 24262, 24264, 24265, 24289, 24321, 24436, 24629, 24631, 24635, 24636, 24638, 24869, 24891, 24892, 24893, 24894, 24896, 24897, 25092, 25095, 25225, 25477 DS50-102-XL-T, Serial Numbers: 24190, 24481, 25467 DS50-103-XL-T, Serial Numbers: 24870, 24898
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2220-2017 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.