Z-2539-2017Class II

Cook Multi-Use Holmium Laser Fibers, Supplied sterilized by ethylene oxide gas in peel-open packages

Cook Incorporated / initiated 2017-05-09 / Terminated / root cause: design

Cook Inc. began a recall of Cook Multi-Use Holmium Laser Fibers, Supplied sterilized by ethylene oxide gas in peel-open packages on . FDA classified it as Class II and gives the reason as "Correction to the product Instructions for Use because the current reprocessing instructions do not provide sufficient detailed information for the cleaning, disinfection, and sterilization of these products". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2539-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Incorporated
Product
Cook Multi-Use Holmium Laser Fibers, Supplied sterilized by ethylene oxide gas in peel-open packages
Product code
GEX Powered Laser Surgical Instrument
Reason for recall
Correction to the product Instructions for Use because the current reprocessing instructions do not provide sufficient detailed information for the cleaning, disinfection, and sterilization of these products.
Root cause tag
design
FDA root cause
Device Design
Action
Cook is initiating a voluntary recall of multiple products in the form of a correction to the product Instructions for Use. Cook Inc. has identified the current reprocessing instructions do not provide sufficient detailed information for the cleaning, disinfection, and sterilization of these products. Potential adverse events that may occur if the products are not adequately reprocessed include urological infections and systemic infections from a urological origin as well as events resulting from chemical residual exposure. Cook distributed notices via courier on May 9, 2017. Customers were instructed to do the following: Distribution of the Cook Multi-Use Holmium Laser Fibers will not occur until the reprocessing instructions in the Instructions for Use have been corrected. Single use fibers are available. Contact your local sales representative for further information. You can continue to use your inventory of Cook Multi-Use Holmium Laser Fibers by following the attached Suggested Fiber Reprocessing Instructions. Action to be taken: 1. Examine your inventory immediately to identify and quarantine those affected products. 2. Implement the updated Suggested Fiber Reprocessing Instructions. 3. Return the required Acknowledgement and Receipt Form within 30 days. 4. Even if you do not have affected products, you must still complete the Acknowledgement and Receipt Form and send it via fax (812.339.7316) or email ([email protected]). 5. Immediately report any adverse events to Cook Medical Customer Relations at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30 am and 5:00 pm (Eastern Time), or by email at [email protected]. Adverse events or quality problems experienced with the use of this product may also be reported to the FDA: Online at: http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail) Call the FDA at: 1-800-FDA-1088 Should you have any medical questions or concer
Quantity in commerce
4369
Distribution
Domestic: AL AR AZ CA CO FL GA HI ID IL IN KY LA MD MI MO MS NC NM NY OH OK OR PA SC TN TX UT VA WA WI Foreign: Austria, Australia, Argentina, Azerbaijan, Belgium, Bulgaria, Bahrain, Canada, Chile, China, Cyprus, Czech Republic, Denmark, Djibouti, Egypt, France, Hong Kong, Honduras, Croatia, Greece, Germany, Hungary, India, Ireland, Italy, Latvia, Malaysia, Morocco, Netherlands, Poland, Portugal, Puerto Rico, Saudi Arabia, Serbia, Spain, Slovenia, South Africa, Sri Lanka, Sweden, Singapore, Taiwan, Thailand, Tunisia, United Kingdom, and Uruguay
Product codes and lots
HLF-M273-H30  HLF-M365-H30  HLF-M550-H30  HLF-M940-H30  HLF-M273-HSMA  HLF-M365-HSMA  HLF-M550-HSMA  HLF-M940-HSMA

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2539-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2539-2017Class IITerminatedVoluntary: Firm initiated