Z-2540-2017Class II

Vue Optic Visualization Source, Flexor Vue Deflecting Endoscopic System, Endoscope, Fiber Optic, Supplied...

Cook Incorporated / initiated 2017-05-03 / Terminated / root cause: design

Cook Inc. began a recall of Vue Optic Visualization Source, Flexor Vue Deflecting Endoscopic System, Endoscope, Fiber Optic, Supplied sterilized by ethylene oxide gas in peel-open packages on . FDA classified it as Class II and gives the reason as "Cook Inc. is initiating a voluntary recall of the Vue Optic Visualization Source and the Flexor Vue Deflecting Endoscopic System because it was determined that the current reprocessing instructions do not provide...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2540-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Incorporated
Product
Vue Optic Visualization Source, Flexor Vue Deflecting Endoscopic System, Endoscope, Fiber Optic, Supplied sterilized by ethylene oxide gas in peel-open packages
Product code
FFS Image, Illumination, Fiberoptic, For Endoscope
Reason for recall
Cook Inc. is initiating a voluntary recall of the Vue Optic Visualization Source and the Flexor Vue Deflecting Endoscopic System because it was determined that the current reprocessing instructions do not provide sufficient detailed information for the cleaning, disinfection, and sterilization of these products.
Root cause tag
design
FDA root cause
Device Design
Action
Cook Inc. initiated a voluntary recall of multiple products because it was identified that the reprocessing instructions do not provide sufficiently detailed information for the cleaning, disinfection, and sterilization of these products. Potential adverse events that may occur if the products are not adequately reprocessed include urological infections and systemic infections from a urological origin as well as events resulting from chemical residual exposure. Notices were mailed on 05/9/2017 via courier. Customers were asked to do the following: Distribution of the VueOptic Visualization Source and Flexor Vue Deflecting Endoscopic System products will not occur until the reprocessing instructions in the Instructions for Use have been corrected. You can continue to use the VueOptic Visualization Source and Flexor Vue Deflecting Endoscopic System by following the attached document for Reprocessing Instructions. Action to be taken: 1. Examine your inventory immediately to identify and quarantine those affected products. 2. Implement the provided updated Reprocessing Instructions. 3. Return the required Acknowledgement and Receipt Form within 30 days. 4. Even if you do not have affected products, you must still complete the Acknowledgement and Receipt Form and send it via fax (812.339.7316) or email ([email protected]). 5. Immediately report any adverse events to Cook Medical Customer Relations at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30 am and 5:00 pm (Eastern Time), or by email at [email protected]. Adverse events or quality problems experienced with the use of this product may also be reported to the FDA: Online at: http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail) Call the FDA at: 1-800-FDA-1088 Should you have any medical questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.
Quantity in commerce
1328 units
Distribution
Domestic: FL IN LA MS NC TX Foreign:Argentina, Australia, Brazil, Canada, Czech Republic, Egypt, Estonia, France, Germany, Israel, Italy, Netherlands, Norway, Paraguay, Poland, Portugal, Russian Federation, Spain, Switzerland, Tunisia, Turkey, United Kingdom, VA/DOD:None
Product codes and lots
Vue Optic Visualization Source, Catalog Number: FVO-150, GPN: G25343 Flexor Vue Deflecting Endoscopic System Catalog Number: FV-090075-150, GPN: G50972 Catalog Number: FV-090045-150, GPN: G34306

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2540-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2540-2017Class IITerminatedVoluntary: Firm initiated