Z-2713-2017Class II

Video Duodenoscope These instruments are intended to provide optical visualization of (via a video monitor)...

PENTAX of America, Inc. / initiated 2017-01-17 / Terminated / root cause: design

Pentax of America Inc began a recall of Video Duodenoscope These instruments are intended to provide optical visualization of (via a video monitor), and therapeutic access to, the biliary tract via... on . FDA classified it as Class II and gives the reason as "Pentax initiated a field correction/safety alert for two (2) models of the Video Duodenoscope to determine how soiling may have occurred on the surface of the suction cylinder and under the distal cap during testing". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2713-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
PENTAX of America, Inc.
Product
Video Duodenoscope These instruments are intended to provide optical visualization of (via a video monitor), and therapeutic access to, the biliary tract via the upper GI tract. This anatomy includes, but is not restricted to, the organs; tissues; and subsystems: esophagus, stomach, duodenum, common bile, hepatic and cystic ducts.
Product code
FDT Duodenoscope And Accessories, Flexible/Rigid
Reason for recall
Pentax initiated a field correction/safety alert for two (2) models of the Video Duodenoscope to determine how soiling may have occurred on the surface of the suction cylinder and under the distal cap during testing.
Root cause tag
design
FDA root cause
Device Design
Action
The firm, Pentax Medical, sent a "FIELD CORRECTION' letter dated January 17, 2017 to their affected customers via USPS Certified Mail. The letter described the product, problem and actions to be taken. The customers were instructed to immediately remove any affected product from use; follow product labeling; ensure all reprocessing personnel are knowledgeable and thoroughly trained on the instructions of Use for manual reprocessing of the devices; clean elevator recesses and follow all reprocessing instructions; and complete and return the FIELD CORRECTION RESPONSE FORM via a member of the PENTAX Service Department as part of an on-site-visit or Fax to: QA/RA Department at 201-799-4063 (alternate 201-391-4189) or a pdf copy to [email protected]. The firm will replace any affected product with its current generation model. If you have questions or request further information or assistance contact PENTAX Medical Customer Support at 1-800-431-5880 (8:30AM - 5:00PM, Monday-Friday, EST) or email: [email protected].
Quantity in commerce
2,015 (US - 519 (ED-3490TK) and 5 (ED-3270K); OUS - 1,491)
Distribution
Worldwide Distribution-US (Nationwide) including states of: AL, AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NH, NJ, NV, NY, OH, OK, PA, PR, TN, TX, UT, VA, WA, WI, WV, WY;. and country of: CANADA.
Product codes and lots
Model Numbers: Ed-3490TK and ED-3270K

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2713-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2713-2017Class IITerminatedVoluntary: Firm initiated