Z-2154-2017Class II
Philips DigitalDiagnost 3.1.x X-Ray System
Philips North America LLC / initiated 2017-05-11 / Terminated / root cause: software
Philips Electronics North America Corporation began a recall of Philips DigitalDiagnost 3.1.x X-Ray System on . FDA classified it as Class II and gives the reason as "During recent evaluations of the Philips DigitalDiagnost 3.1.x, we have identified a potential issue that may affect the performance of the equipment under certain conditions". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2154-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips DigitalDiagnost 3.1.x X-Ray System
- Product code
- MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
- Reason for recall
- During recent evaluations of the Philips DigitalDiagnost 3.1.x, we have identified a potential issue that may affect the performance of the equipment under certain conditions.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Philip's planned action to bring defect into compliance: 1. You will contact customers and initiate a software update to correct the issue that incorrectly notifies the user of the detectors being ready. 2. You will install the software/firmware and perform testing to ensure the software update was effective. 3.You will provide a customer notification letter that includes a statement that you will without charge, remedy the defect or bring the product into compliance. CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 2.Philips Healthcare will implement this CAP by November 10, 2017. If you need any further information or support concerning this Issue, please contact your local Philips representative: Technical Support Line: 1-800-722-9377. We apologize for any inconvenience this may cause and trust that this information is adequately addressing any concerns you may have.
- Quantity in commerce
- US - 132
- Distribution
- USA(nationwide) Distribution to the states of : AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KY, LA, MA, ME, MI, MN, MO, NC, ND, NE, NH, NM, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA and WI
- Product codes and lots
- DigitalDiagnost Upgrade Release 3.1.x to 3.1.3
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2154-2017 | Class II | Terminated | FDA Mandated |