510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2688-2017Class II

Sensis Vibe Systems with Software Version VD10B, Model Numbers 10765502, 10910620, 11007641, 6648161...

Siemens Medical Solutions USA, Inc. / initiated 2017-05-16 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Sensis Vibe Systems with Software Version VD10B, Model Numbers 10765502, 10910620, 11007641, 6648161 --- Programmable diagnostic computer on . FDA classified it as Class II and gives the reason as "Software error. In Sensis Vibe systems with software version VD10B, a software error can result in: problems generating a report and/ or - information from different examinations of the same patient being combined into...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2688-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Sensis Vibe Systems with Software Version VD10B, Model Numbers 10765502, 10910620, 11007641, 6648161 --- Programmable diagnostic computer,
Product code
DQK Computer, Diagnostic, Programmable
Reason for recall
Software error. In Sensis Vibe systems with software version VD10B, a software error can result in: problems generating a report and/ or - information from different examinations of the same patient being combined into one report. --- The error causes information from two examinations to be combined into one report. This does not result in a hazardous situation; however, if the erroneous report has been used for treatment or diagnosis, this can potentially result in an incorrect treatment or diagnosis.
Root cause tag
software
FDA root cause
Software design
Action
Siemens sent an Customer Safety Advisory Notice dated May 16, 2017, was mailed to all users on May 16, 2017. The letter informs customers of the software issue and potential impact. Siemens Medical Solutions will contact users to make arrangements to update the system with updated software to correct the issue. For further questions, please call (610) 448-6461.
Quantity in commerce
US: 13 units
Distribution
US Distribution to the states of : IN, MI, IA
Product codes and lots
Software Version VD10B

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2688-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2688-2017Class IITerminatedVoluntary: Firm initiated