Z-2761-2017Class II

Atrium iCAST Balloon Expandable Covered Stent; 7mm x 22mm x 120cm; Model/Part Number 85455 (Marketed in U.S...

Atrium Medical Corporation / initiated 2017-05-25 / Terminated / root cause: packaging

Atrium Medical Corporation began a recall of Atrium iCAST Balloon Expandable Covered Stent; 7mm x 22mm x 120cm; Model/Part Number 85455 (Marketed in U.S. only) Atrium Advanta V12 Balloon Expandable... on . FDA classified it as Class II and gives the reason as "Firm received a complaint that a customer who ordered an ADVANTA V12 7mm x 22mm x 120cm received an ADVANTA v12 7 mm x 38 mm x 80 cm, due to the manifold (catheter hub) was printed as 7mm x 38 mm x 80 cm. The complaint...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2761-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Atrium Medical Corporation
Product
Atrium iCAST Balloon Expandable Covered Stent; 7mm x 22mm x 120cm; Model/Part Number 85455 (Marketed in U.S. only) Atrium Advanta V12 Balloon Expandable Covered Stent; 7mm x 22mm x 120cm; Model/Part Number 85355 (Marketed OUTSIDE U.S. only) The iCAST Covered Stent is indicated for the treatment of tracheobronchial strictures produced by malignant neoplasms.
Product code
JCT Prosthesis, Tracheal, Expandable
Reason for recall
Firm received a complaint that a customer who ordered an ADVANTA V12 7mm x 22mm x 120cm received an ADVANTA v12 7 mm x 38 mm x 80 cm, due to the manifold (catheter hub) was printed as 7mm x 38 mm x 80 cm. The complaint was the result of incorrect dimensional information (7mm x 38mm x 80cm) printed on the manifiold (catheter hub). The complaint device physical dimensions matched the dimensional information displayed on the outer packaging and the device inner pouch. This was the only complaint received for this non-conformance.
Root cause tag
packaging
FDA root cause
Under Investigation by firm
Action
The firm, Getinge/Maquet, sent an "URGENT MEDICAL DEVICE RECALL NOTICE" dated May 25, 2017 via FedEx overnight to affected US consignees via FedEx Next Day Air. Getinge sent a second "URGENT MEDICAL DEVICE RECALL LETTER" dated June 2, 2017 via FedEx overnight to the affected customers in order to correct a product code/part number due to a typo in the initial letter and to assist in identifying the labeling problem. The letters described the product, problem and actions to be taken. Customers are instructed to examine your inventory; remove affected products; quarantine and place in a secure location; if you have the affected product, contact Getinge Customer Service at 1-603-880-1433, Monday through Friday between 9:00 am to 5:00 pm for RA and shipping instructions to return affected product to the Manufacturer, and complete and Return the enclosed Medical Device - Recall Response Form to email: [email protected] or fax to 1-973-807-9207. If you have any questions, please contact your Getinge/Maquet representative or call Maquet Customer Service at 1-603-880-1433, Monday through Friday between the hours of 9:00 am to 5:00 pm Eastern Time Zone.
Quantity in commerce
246 (US: 183 units; Foreign: 63 units)
Distribution
Worldwide distribution-US (nationwide) and countries of: Germany, Great Britain, Austria, Ireland, Netherlands, France, Italy, Sweden, and San Marino.
Product codes and lots
Part Number 85455: Lots 240735, 240523;  Part Number 85355: Lot 240734

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2761-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2761-2017Class IITerminatedVoluntary: Firm initiated