Z-2746-2017Class II

Install CD, Spine tools, Plus and S7, Version 25. Model Number 9731958

Medtronic Navigation, Inc / initiated 2017-05-10 / Terminated / root cause: software

Medtronic Navigation, Inc. began a recall of Install CD, Spine tools, Plus and S7, Version 25. Model Number 9731958 on . FDA classified it as Class II and gives the reason as "Software issue related to the StealthStation S7 system and the Synergy Spine application Version 2.1 configured with Spine Tool Install CD version 25. Issue may result in user being unable to navigate the Navigated...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2746-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
Install CD, Spine tools, Plus and S7, Version 25. Model Number 9731958.
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
Software issue related to the StealthStation S7 system and the Synergy Spine application Version 2.1 configured with Spine Tool Install CD version 25. Issue may result in user being unable to navigate the Navigated Elevate Inserter with the StealthAiR Spine Frame during spine surgical procedures.
Root cause tag
software
FDA root cause
Software in the Use Environment
Action
Medtronic sent an Urgent - Product Advisory Notice dated May 10, 2017, to all affected customers. Instructions indicated that a Medtronic representative will be in contact to update the system with version 26 and will remove any copies of the Spine Tools Install CD, version 25. Please quarantine any copies of version 25 until they can be removed. If you have further questions regarding this communication, please contact your local Medtronic Navigation Technical Services Representative, your local Medtronic Spine Sales Representative, or call (800 595-9709.
Quantity in commerce
108
Distribution
Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Brazil, Germany, Switzerland and Great Britain.
Product codes and lots
Kits that include affected 9731958 INSTALL CD: INST SET 9734833 UNIVERSAL NAVLOCK, INST SET 9734834 CAPSTONE TRIALS, INST SET 9734835 CLYDESDALE TRIALS, INST SET 9734836 TLIF/DLIF INSERTER, INST SET 9735283 SOLERA 5.5/6.0 DRIVERS, INST SET 9735284 SOLERA NONCAN TAPS, INST SET 9735285 SOLERA ILIAC TAPS, TAPS KIT 9735465 SOLERA AWL-TIP TAPS, REF SET 9735500 SPINE REFRENCING, UPGRD KIT 9735522 SPINE REVISION, INSTR SET 9735607 VERTEX SELECT

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2746-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2746-2017Class IITerminatedVoluntary: Firm initiated