Z-0171-2018Class II

BD Phoenix PMIC-107, Catalog Number 448417, UDI #s (01)30382904484171, (10)7031818(17)180228(30)0025, IVD...

Becton Dickinson & Co / initiated 2017-05-11 / Terminated / root cause: manufacturing process

Becton Dickinson & Co. began a recall of BD Phoenix PMIC-107, Catalog Number 448417, UDI #s (01)30382904484171, (10)7031818(17)180228(30)0025, IVD. PMIC-107 is a Gram positive AST Only panel used to... on . FDA classified it as Class II and gives the reason as "BD has recently confirmed quality control failures, related to a manufacturing issue, for the inducible macrolide-linosamide-streptogramin B (iMLSb) test contained on lot 7031818 of Phoenix panels. The iMLSb resistance...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0171-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
BD Phoenix PMIC-107, Catalog Number 448417, UDI #s (01)30382904484171, (10)7031818(17)180228(30)0025, IVD. PMIC-107 is a Gram positive AST Only panel used to test antibiotic susceptibility. The inducible macrolide-linosamide-steptogramin B (iMLSb) resistance D-test screen for Staphylococcus species is used to identify phenotypic drug resistance to clindamycin and erythromycin. --- Product is packaged 25 panels to 1 box.
Product code
LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
Reason for recall
BD has recently confirmed quality control failures, related to a manufacturing issue, for the inducible macrolide-linosamide-streptogramin B (iMLSb) test contained on lot 7031818 of Phoenix panels. The iMLSb resistance D-test screen for Staphylococcus species is used to identify phenotypic drug resistance to clindamycin and erythromycin. As the product should not be used when quality control (QC) fails, it is unlikely that there would be any impact on patient results, however, if a customer does not perform QC testing and uses the panel, there is a potential to incorrectly treat a Staphylococcus infection with clindamycin based on the iMLSb test result. This issue can be detected 100% of the time when used per package insert instructions with iMLSb positive strain S. aureus BAA-977.
Root cause tag
manufacturing process
FDA root cause
Manufacturing material removal
Action
URGENT PRODUCT RECALL LETTER (dated May 2017) were sent to customers via a UPS mailing on May 11, 2017. Customers were instructed to discontinue use of the affected lot and discard any remaining product. BD will replace the discarded inventory. Please complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification and process your product replacement. Please complete and fax the form to: Attention: Regulatory Compliance, Fax: 410-316-4258 or e-mail to: [email protected]. For further assistance, please contact BD (Customer/Technical) Support at 1-800-638-8663 in the USA.
Quantity in commerce
US: 7750 units
Distribution
CT, VA, PA, TX, FL, NC, WI, NM
Product codes and lots
Lot 7031818, Exp. Date 2/28/2018

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0171-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0171-2018Class IITerminatedVoluntary: Firm initiated