Z-2770-2017Class II
VITEK 2 Gram Negative test kits containing colistin (cs01n)
bioMerieux, Inc. / initiated 2017-05-16 / Terminated
Biomerieux Inc began a recall of VITEK 2 Gram Negative test kits containing colistin (cs01n) on . FDA classified it as Class II and gives the reason as "The investigation demonstrated a high rate of very major errors (resistant isolates calling susceptible) with the VITEK 2 AST-GN colistin (cs01n) compared to agar dilution (the reference method used for cs01n...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2770-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- bioMerieux, Inc.
- Product
- VITEK 2 Gram Negative test kits containing colistin (cs01n)
- Product code
- PTJ Phenotypic Test Kit, Non-Susceptible/Elevated Mic Organisms, Cultured Isolates
- Reason for recall
- The investigation demonstrated a high rate of very major errors (resistant isolates calling susceptible) with the VITEK 2 AST-GN colistin (cs01n) compared to agar dilution (the reference method used for cs01n development) and compared to broth-microdilution (recommended method by EUCAST/CLSI).
- FDA root cause
- Other
- Action
- Customers were notified on or about 05/16/2017 and then notified again with an updated letter. Instructions to customers included requesting that the attached acknowledgement form be completed, distribute the recall notification to customers, and provide the notice to future customers.
- Quantity in commerce
- 1,520,000 cards
- Distribution
- Domestic: NC. Foreign: Argentina, Austria, Brazil, Canada, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, France, Germany, Guatemala, Hong Kong, India, Israel, Italy, Japan, Malaysia, Mexico, Netherlands, Netherlands Antilles, Nicaragua, Panama, Paraguay, Peru, Philippines, Poland, Singapore, South Africa, South Korea, Spain, Thailand, Turkey, United Kingdom, Uruguay, Venezuela, & Vietnam.
- Product codes and lots
- REF # (Description): 418114 (AST-GN94), 22258 (AST-N102), 412605 (AST-N194), 412921 (AST-N199), 412862 (AST-N201), 412863 (AST-N202), 412924 (AST-N205), 412918 (AST-N210), 413061 (AST-N212), 413083 (AST-N222), 413112 (AST-N225), 413171 (AST-N236), 413172 (AST-N237), 413204 (AST-N239), 413205 (AST-N240), 413391 (AST-N242), 413394 (AST-N245), 413397 (AST-N248), 413572 (AST-N249), 413722 (AST-N253), 413724 (AST-N255), 413725 (AST-N256), 413753 (AST-N261), 414164 (AST-N272), 414286 (AST-N276), 414492 (AST-N279), 414531 (AST-N280), 414532 (AST-N281), 414972 (AST-N288), 415063 (AST-N292), 415369 (AST-N293), 415433 (AST-N295), 416005 (AST-N299), 416241 (AST-N300), 416590 (AST-N303), 416913 (AST-N308), 416914 (AST-N309), 417426 (AST-N315), 417952 (AST-N318), 418204 (AST-N320), 418489 (AST-N322), 418513 (AST-N235), 418514 (AST-N326), 418630 (AST-N328), 418631 (AST-N329), 418675 (AST-N331), 418676 (AST-N332), 418985 (AST-N335), 419077 (AST-N338), 419341 (AST-N339), 420063 (AST-N343), 420440 (AST-N344), 420867 (AST-N347), 420856 (AST-N348), 421258 (AST-N352), 421297 (AST-N353), 421350 (AST-N354), 421351 (AST-N355, 421451 (AST-N357), 421441 (AST-N358), 421583 (AST-N360), 421584 (AST-N361), 421585 (AST-N362), 421692 (AST-363), 421693 (AST-N364), 421830 (AST-N365), 421853 (AST-N366), 421854 (AST-N367), 421855 (AST-N368), 410025 (AST-XN01), 413230 (AST-XN05)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2770-2017 | Class II | Terminated | Voluntary: Firm initiated |