Z-2770-2017Class II

VITEK 2 Gram Negative test kits containing colistin (cs01n)

bioMerieux, Inc. / initiated 2017-05-16 / Terminated

Biomerieux Inc began a recall of VITEK 2 Gram Negative test kits containing colistin (cs01n) on . FDA classified it as Class II and gives the reason as "The investigation demonstrated a high rate of very major errors (resistant isolates calling susceptible) with the VITEK 2 AST-GN colistin (cs01n) compared to agar dilution (the reference method used for cs01n...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2770-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
bioMerieux, Inc.
Product
VITEK 2 Gram Negative test kits containing colistin (cs01n)
Product code
PTJ Phenotypic Test Kit, Non-Susceptible/Elevated Mic Organisms, Cultured Isolates
Reason for recall
The investigation demonstrated a high rate of very major errors (resistant isolates calling susceptible) with the VITEK 2 AST-GN colistin (cs01n) compared to agar dilution (the reference method used for cs01n development) and compared to broth-microdilution (recommended method by EUCAST/CLSI).
FDA root cause
Other
Action
Customers were notified on or about 05/16/2017 and then notified again with an updated letter. Instructions to customers included requesting that the attached acknowledgement form be completed, distribute the recall notification to customers, and provide the notice to future customers.
Quantity in commerce
1,520,000 cards
Distribution
Domestic: NC. Foreign: Argentina, Austria, Brazil, Canada, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, France, Germany, Guatemala, Hong Kong, India, Israel, Italy, Japan, Malaysia, Mexico, Netherlands, Netherlands Antilles, Nicaragua, Panama, Paraguay, Peru, Philippines, Poland, Singapore, South Africa, South Korea, Spain, Thailand, Turkey, United Kingdom, Uruguay, Venezuela, & Vietnam.
Product codes and lots
REF # (Description): 418114 (AST-GN94), 22258 (AST-N102), 412605 (AST-N194), 412921 (AST-N199), 412862 (AST-N201), 412863 (AST-N202), 412924 (AST-N205), 412918 (AST-N210), 413061 (AST-N212), 413083 (AST-N222), 413112 (AST-N225), 413171 (AST-N236), 413172 (AST-N237), 413204 (AST-N239), 413205 (AST-N240), 413391 (AST-N242), 413394 (AST-N245), 413397 (AST-N248), 413572 (AST-N249), 413722 (AST-N253), 413724 (AST-N255), 413725 (AST-N256), 413753 (AST-N261), 414164 (AST-N272), 414286 (AST-N276), 414492 (AST-N279), 414531 (AST-N280), 414532 (AST-N281), 414972 (AST-N288), 415063 (AST-N292), 415369 (AST-N293), 415433 (AST-N295), 416005 (AST-N299), 416241 (AST-N300), 416590 (AST-N303), 416913 (AST-N308), 416914 (AST-N309), 417426 (AST-N315), 417952 (AST-N318), 418204 (AST-N320), 418489 (AST-N322), 418513 (AST-N235), 418514 (AST-N326), 418630 (AST-N328), 418631 (AST-N329), 418675 (AST-N331), 418676 (AST-N332), 418985 (AST-N335), 419077 (AST-N338), 419341 (AST-N339), 420063 (AST-N343), 420440 (AST-N344), 420867 (AST-N347), 420856 (AST-N348), 421258 (AST-N352), 421297 (AST-N353), 421350 (AST-N354), 421351 (AST-N355, 421451 (AST-N357), 421441 (AST-N358), 421583 (AST-N360), 421584 (AST-N361), 421585 (AST-N362), 421692 (AST-363), 421693 (AST-N364), 421830 (AST-N365), 421853 (AST-N366), 421854 (AST-N367), 421855 (AST-N368), 410025 (AST-XN01), 413230 (AST-XN05)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2770-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2770-2017Class IITerminatedVoluntary: Firm initiated