Z-2193-2017Class II
GE Healthcare Discovery Ml X-ray system
GE Healthcare / initiated 2017-05-19 / Terminated / root cause: software
GE Healthcare, LLC began a recall of GE Healthcare Discovery Ml X-ray system on . FDA classified it as Class II and gives the reason as "It was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2193-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE Healthcare Discovery Ml X-ray system
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- It was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- GE Healthcare proposed action to bring defect into compliance : 1. Inform users of the potential for this issue to occur and instruct users as to the actions to take in the event of an Emergency Stop actuation during prep-delay. 2. Provide an update to product labeling for users that includes this information. GE Healthcare will include the updated product labeling with all current production systems. 3. Make modifications during the next scheduled product update to prevent this "race" condition from occurring in newly produced systems. CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 2. Any future customer notification letters to be sent must comply with the requirements of 21 CFR 1003.21. This includes, but is not limited to, the prescribed packaging and the statement that you will remedy the defect or bring the product into compliance at no charge. 3.You complete this recall by September 1, 2017. Report any failures regarding this recall immediately to the District Recall Coordinator. For further questions, please call (262) 513-4122.
- Quantity in commerce
- 10,884
- Distribution
- Worldwide Distribution - USA (nationwide) including Puerto Rico and to the countries of : Japan, Jordan, Kazakhstan, Korea ( Republic of) Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mauritania, Mexico, Moldova, Morocco, Nambia, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic of Serbia, Reunion, Romania, Russia, Rwanda, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain Sweden, Switzerland, Syria, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietam and Yemen.
- Product codes and lots
- n/a
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2181-2017 | Class II | Terminated | FDA Mandated | ||
| Z-2182-2017 | Class II | Terminated | FDA Mandated | ||
| Z-2183-2017 | Class II | Terminated | FDA Mandated | ||
| Z-2184-2017 | Class II | Terminated | FDA Mandated | ||
| Z-2185-2017 | Class II | Terminated | FDA Mandated | ||
| Z-2186-2017 | Class II | Terminated | FDA Mandated | ||
| Z-2187-2017 | Class II | Terminated | FDA Mandated | ||
| Z-2188-2017 | Class II | Terminated | FDA Mandated | ||
| Z-2189-2017 | Class II | Terminated | FDA Mandated | ||
| Z-2190-2017 | Class II | Terminated | FDA Mandated | ||
| Z-2191-2017 | Class II | Terminated | FDA Mandated | ||
| Z-2192-2017 | Class II | Terminated | FDA Mandated | ||
| Z-2193-2017 | Class II | Terminated | FDA Mandated | ||
| Z-2194-2017 | Class II | Terminated | FDA Mandated | ||
| Z-2195-2017 | Class II | Terminated | FDA Mandated | ||
| Z-2196-2017 | Class II | Terminated | FDA Mandated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.