Z-3067-2017Class II

Ultrasound Video Gastroscope, Model EG-3670URK, The product is intended to be used in combination with the...

PENTAX of America, Inc. / initiated 2017-05-23 / Terminated / root cause: labeling and UDI

Pentax of America Inc began a recall of Ultrasound Video Gastroscope, Model EG-3670URK, The product is intended to be used in combination with the ultrasound scanning unit made by Hitachi Medical... on . FDA classified it as Class II and gives the reason as "Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the cleaning detergent...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-3067-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
PENTAX of America, Inc.
Product
Ultrasound Video Gastroscope, Model EG-3670URK, The product is intended to be used in combination with the ultrasound scanning unit made by Hitachi Medical Corporation, and the PENTAX Medical video processor to provide images for observation, diagnosis, imaging, or therapy of the upper gastrointestinal tract from esophagus through duodenum.
Product code
ODG Endoscopic Ultrasound System, Gastroenterology-Urology
Reason for recall
Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the cleaning detergent solution should remain in contact with the all internal channels and external endoscope surfaces. The disinfecting solution, not the cleaning detergent, should be used for this step.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
Recall Notifications were sent to 104 consignees via USPS Certified Mail on 5/23/2017. Included with the Field Safety letter was the customer response form, and the addendum to IFU Z845. These were also posted on Pentax of America web page until the recall is closed.
Quantity in commerce
US: 118 units
Distribution
Nationwide
Product codes and lots
Serial Numbers: G110307, H110482, H110484, H110487, H110489, G110334, G110338, G110404, G110296, G110358, G110340, G110341, G110407, G110409, H110432, G110345, H110450, G110391, H110476, H110443, H110446, G110400, H110483, G110394, G110406, H110411,  H110497, G110360, G110363, G110364, G110371, G110366, G110368,  G110372, H110434, H110447, G110380, H110455, G110298, H110445,  H110449, H110461, G110292, G110346, G110365, H110494, H110495,  H110474, G110376, G110319, G110328, H110473, H110413, H110414, H110427, H110453, H110454, H110415, G110312, G110321, H110467,  H110471, H110412, G110289, G110303, G110294, H110420, G110299,  G110300, H110448, H110452, G110386, G110331, H110451, H110493,  G110313, G110314, H110464, H110496, H110433, G110373, G110375,  G110317, H110431, H110417, G110291, G110302, G110369, G110378,  G110335, G110387, G110390, G110392, G110395, G110311, G110327,  G110332, G110382, G110355, G110357, G110287, G110320, H110435,  H110481, H110485, H110486, G110354, G110316, G110318, G110362,  G110393, H110498, H110456, H110457, H110459, G110337, G110356,  H110421, H110429, G110290, G110352, H110470, H110478, G110383,  G110384, G110361.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3067-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3067-2017Class IITerminatedVoluntary: Firm initiated
Z-3068-2017Class IITerminatedVoluntary: Firm initiated