Ultrasound Video Gastroscope, Model EG-3670URK, The product is intended to be used in combination with the...
Pentax of America Inc began a recall of Ultrasound Video Gastroscope, Model EG-3670URK, The product is intended to be used in combination with the ultrasound scanning unit made by Hitachi Medical... on . FDA classified it as Class II and gives the reason as "Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the cleaning detergent...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-3067-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- PENTAX of America, Inc.
- Product
- Ultrasound Video Gastroscope, Model EG-3670URK, The product is intended to be used in combination with the ultrasound scanning unit made by Hitachi Medical Corporation, and the PENTAX Medical video processor to provide images for observation, diagnosis, imaging, or therapy of the upper gastrointestinal tract from esophagus through duodenum.
- Product code
- ODG Endoscopic Ultrasound System, Gastroenterology-Urology
- Reason for recall
- Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the cleaning detergent solution should remain in contact with the all internal channels and external endoscope surfaces. The disinfecting solution, not the cleaning detergent, should be used for this step.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Recall Notifications were sent to 104 consignees via USPS Certified Mail on 5/23/2017. Included with the Field Safety letter was the customer response form, and the addendum to IFU Z845. These were also posted on Pentax of America web page until the recall is closed.
- Quantity in commerce
- US: 118 units
- Distribution
- Nationwide
- Product codes and lots
- Serial Numbers: G110307, H110482, H110484, H110487, H110489, G110334, G110338, G110404, G110296, G110358, G110340, G110341, G110407, G110409, H110432, G110345, H110450, G110391, H110476, H110443, H110446, G110400, H110483, G110394, G110406, H110411, H110497, G110360, G110363, G110364, G110371, G110366, G110368, G110372, H110434, H110447, G110380, H110455, G110298, H110445, H110449, H110461, G110292, G110346, G110365, H110494, H110495, H110474, G110376, G110319, G110328, H110473, H110413, H110414, H110427, H110453, H110454, H110415, G110312, G110321, H110467, H110471, H110412, G110289, G110303, G110294, H110420, G110299, G110300, H110448, H110452, G110386, G110331, H110451, H110493, G110313, G110314, H110464, H110496, H110433, G110373, G110375, G110317, H110431, H110417, G110291, G110302, G110369, G110378, G110335, G110387, G110390, G110392, G110395, G110311, G110327, G110332, G110382, G110355, G110357, G110287, G110320, H110435, H110481, H110485, H110486, G110354, G110316, G110318, G110362, G110393, H110498, H110456, H110457, H110459, G110337, G110356, H110421, H110429, G110290, G110352, H110470, H110478, G110383, G110384, G110361.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3067-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3068-2017 | Class II | Terminated | Voluntary: Firm initiated |