TI Transport Monitor/Module Battery Pack
Mindray DS USA, Inc. dba Mindray North America began a recall of TI Transport Monitor/Module Battery Pack Product Usage: intended for monitoring, displaying, reviewing, storing, alarming and transferring of multiple... on . FDA classified it as Class II and gives the reason as "Mindray has received two reports from outside the United States that the battery pack for TI transport monitor/module overheated". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2634-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Datascope Corp.
- Product
- TI Transport Monitor/Module Battery Pack Product Usage: intended for monitoring, displaying, reviewing, storing, alarming and transferring of multiple physiological parameters.
- Product code
- MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Reason for recall
- Mindray has received two reports from outside the United States that the battery pack for TI transport monitor/module overheated.
- FDA root cause
- Under Investigation by firm
- Action
- Mindray notified their customers on April 28, 2017.Customers are provided information regarding this correction. Once the customer has identified the batteries, the customer should contact Mindray Customer Service to arrange for the no charge delivery of replacements. Mindray Customer Service representatives may be reached at (800) 288-2121 (Option 3), Monday through Friday 8:30 amd -5:30 pm, ET.
- Quantity in commerce
- 2718 batteries
- Distribution
- US Nationwide Distribution
- Product codes and lots
- two sales brochures: p/n: 0002-08-30034 and 0002-08-30035 two copies of labels: 047-014270-00, 047-010964-00, 047-018754-00 two copies of operator's manuals: 115-040812-00 and TNC 14-FZ007-F-001
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2634-2017 | Class II | Terminated | Voluntary: Firm initiated |